* Signal sourcing unit range and accuracy
DC voltage:-10–110 mV; ±(0.02% + 0.1 mV) at 0–1.1 V
DC current: 0–24 mA, 4/8/12/16/20 mA; ±(0.025% + 3 µA)
mA SINK: 0.1–24 mA; ±(0.05% + 3 µA)
Resistance: 0–400.00 Ω; ±(0.025% + 0.1 Ω)
RTD : -200–850°C; ±(0.025% + 0.3°C)
TC: + K: -200–1372°C; ±(0.02% + 0.5°C) (-100°C or greater)
+ E: -200–1000°C; ±(0.02% + 0.5°C) (-100°C or greater)
+ J: -200–1200°C ; ±(0.02% + 0.5°C) (-100°C or greater)
+ T: -200.0–400.0°C; ±(0.02% + 0.5°C) (0°C or greater)
+ N: -200–1300°C; ±(0.02% + 0.5°C) (0°C or greater)
+ L: -200–900°C; ±(0.02% + 0.5°C) (0°C or greater)
+ U: -200.0–400.0°C;±(0.02% + 0.5°C) (0°C or greater)
+ R: 0–1768°C; ±(0.02% + 1.5°C) (100°C or greater)
+ S: 0–1768°C; ±(0.02% + 1.5°C) (100°C or greater)
+ B: 600–1800°C; ±(0.02% + 1.5°C) (1000°C or greater)
Frequency, pulse: 1–500 Hz, 90–1100 Hz , 0.9 kHz–11 kHz , 1–99,999 cycles; ±0.1 kHz at 0.9 kHz–11 kHz
* Measurement unit range and accuracy
DC voltage: 0–±110 mV, 0–±1.1 V, 0–±11.000 V, 0–±110.00 V; ±(0.025% + 0.2 mV) at 0–±1.1000 V
DC current: 0–±24 mA, 0–±100 mA; ±(0.025% + 4 µA) at 0–±24 mA
Resistance: 0–400 Ω; ±(0.05% + 0.1 Ω)
AC voltage: 0–1.1V, 0–11V, 0–110V, 0–300 V; ±(0.5% + 2 dgt) at 0–300 V
Frequency,pulse: 1–100 Hz , 1.0–1000Hz , 0.001–11 kHz, 0–99,999 CPM, 0–99,999 CPH; ±2 dgt
* Measurement unit (temperature) range and accuracy
TC: + K: -200.0–1372.0°C; ±(0.05% + 1.5°C) (-100°C or greater)
+ E: -200.0–1000.0°C ; ±(0.05% + 1.5°C) (-100°C or greater)
+ J: -200.0–1200.0°C ; ±(0.05% + 1.5°C) (-100°C or greater)
+ T: -200.0–400.0°C ; ±(0.05% + 1.5°C) (-100°C or greater)
+ N: -200.0–1300.0°C ; ±(0.05% + 1.5°C) (-100°C or greater)
+ L: -200.0–900.0°C; ±(0.05% + 1.5°C) (-100°C or greater)
+ U: -200.0–400.0°C ; ±(0.05% + 1.5°C) (-100°C or greater)
+ R: 0–1768°C ; ±(0.05% + 2°C) (100°C or greater)
+ S: 0–1768°C ; ±(0.05% + 2°C) (100°C or greater)
+ B: 600–1800°C ; ±(0.05% + 2°C) (100°C or greater)
RTD : -200.0–850.0°C , -200.0–500.0°C; ±(0.05% + 0.6°C)
Standard accessories: Signal generating lead cables (one red, two black): 98020, Measurement lead cables (one red, one black): RD031, Carrying case: 93016, Terminal adapter for CA71: 99021, User’s manual: IM CA71-E Fuse: 99040 (for current terminal input protection), Four AA alkaline batteries: A1070EB × 4
Optional accessories (sold separately): AC adapter: 94013 (120 V AC power supply), RJ sensor: B9108WA (For reference junction compensation), Accessory carrying case: B9108XA, Communication cable: 91017