Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Supplied output range: 0 to 200 kPa gauge
Minimum set resolution: 0.01 kPa
Supplied output: 280 ± 20kPa
Max. allowable input: 500 kPa gauge
Including calibration accuracy: 60.05% of full scale (at 23°C ±3°C)
Not including calibration accuracy: 60.045% of full scale (at 23°C ±3°C)
Output noise: 60.02% of full scale
Effect of mounting orientation: Forward/backward incline of 90°: ±0.01% of full scale/ Sideways incline of 30°: ±0.2% of full scale
Temperature coefficient: Zero point: ±0.003% of full scale/ °C/ Span: ±0.002% of full scale/ °C
Pressure display unit: kPa only; kPa, kgf/cm2, mmHg, mmH2O (selectable); kPa, inH2O, inHg, psi (selectable)
Output settings: 4.5-digit settings
Alarm : LED turns on for low or excessively high supply pressure.
Supply pressure source : Dry air only: Temperature must be between
5°C and 40°C, and the amount of temperature change must be small.
A pressure-reducing valve with a filter must be used to input a stable supply pressure.
Pressure sensor: Silicon resonant sensor
Output response time (Time for value to read 60.1% of full scale once
change starts): Approximately 5 seconds/ Conditions: Any 20% - or 25% - divided output (one step), with no load.
Monitor output: 0 to 10 mV/full scale or 0 to 2 V/full scale (selectable)
Calibration interval: Six months
Air consumption rate: Approximately 30 liters per minute (with supply pressure in specified range)
Warmup time: Approximately 5 minutes
Operating temperature and humidity ranges: 5 to 40°C and 20 to 80% RH (no condensation)
Maximum operating altitude: 2000 meters
Storage temperature range: -20 to 608C
AC power ratings: 100-120/200-240 V AC, 50/60 Hz
Power fluctuation tolerance range: 90-132 V AC/180-264 V AC
Frequency fluctuation tolerance range: 47-63 Hz
Power consumption: 40 VA Max. (100-200V) / 50 VA Max. (200-240V)
Insulation resistance: Minimum 100 MΩ at 500 V DC (across AC
power and casing)
Withstand voltage: 1500 V AC, 50/60 Hz, for one minute (across AC
power and casing)
Dimensions: Approximately 132 x 213 x 400 mm (protrusions not included)
Weight: approximately 9.5 kg
Accessories: Input adapter connectors (For Φ4 x Φ6 PVC tube, B9310RR), Two rubber pads for rear feet, one power cord, Fuse (A1113EF), one instruction manual.