For full functionality of this site it is necessary to enable JavaScript.
EMIN.CO.TH
0
Product image

Tunix TCMTHC Temperature & Humidity Calibration Chamber (0°C ~ 50°C, 20% RH ~ 95% RH)

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Hi performance Temperature / Humidity Calibration chamber with Top glass opening , HMI Touch panel

Temperature

Temperature Range: 0°C to 50°C

Thermal Stability @50 %RH: ±0.2°C @0°C,@25°C;  ±0.8°C @50°C

Come Up Time (Time Required to achieve target temperature from ambient temperature): 45 Minutes @0°C; 20 Minutes @50°C

Stabilization Time: 15 Minutes 

Thermal Non uniformity @50 % RH: ± 1.5°C

Relative Humidity

Relative Humidity Range @25°C: 20% RH to 95% RH

Relative Humidity  Stability @25°C: ±0.5%RH  @20%RH @50%RH; ±2.2%RH @95%RH

Come Up Time (Time Required to achieve target Relative Humidity from @50 %RH):  30 Minutes @20 %RH; 20 Minutes @ 95%RH

Stabilization Time: 15 Minutes

Relative Humidity Non uniformity @25°C:  ±1.5%RH @20%RH @50%RH; ±3%RH @95%RH

Temperature rise from ambient to maximum value @ambient of 25°C (25°C~50°C): NMT 20 Minutes

Temperature fall from ambient to minimum value @ambient of 25°C (25°C~0°C): NMT 45 Minutes

Temperature rise from Minimum to maximum value @ambient of 25°C (0°C~50°C): NMT 25 Minutes

Temperature fall from Maximum to minimum value @ambient of 25°C (50°C~0°C): NMT 60 Minutes

Datasheet

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts