For full functionality of this site it is necessary to enable JavaScript.
EMIN.CO.TH
0
Product image

TOB TOB-GB-800S Lab Glove Box (+/-10 ~ +/-12mbar)

ModelTOB-GB-800S
OriginChina
Warranty12 months
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Glove box

Box

Box size: (L*W*H)=1220×750×900mm

Box material: stainless steel 304, thickness 3mm

Inner surface:Stainless steel oil film drawing treatment

Outside surface: paint, white

Front window: Tilt windows, transparent tempered safety glass, thickness 8 mm

Glove ports: The material isaluminum, O-ring seal, self-lubricating performance, easy to maintain, and good sealing performance

Gloves: Butyl rubber, diameter 8", length 32", thickness≥0.4mm

Filter:0.3 microns, a gas inlet and a gas outlet

Shelf: Stainless steel, 2 layers, height adjustable

Lighting: LED lights, installed in front of each glass window

Connector: Standby interface 3pcs, DN 40 KF, Power interface 1pc (220V)

BiggerAntechamber

Chamber

Size: diameter 360mm, length 600mm

Material: 304 stainless steel

Accessory

Sliding tray: 304 stainless steel

Door: double door, anodized alumina material, thickness 10mm, vertical operation, with lifting mechanism

Pressure gauge: Analog display

Control: Solenoid valve touch screen automatic operation

SmallerAntechamber

Chamber

Size: diameter 150mm, length 300mm (100mm length into the glove box)

Material: 304 stainless steel

Surface: inner: wireoil filmdrawingtreatment, outer: painting (white)

Accessory

Tray:304 stainless steel

Door: double door,Button flip

Pressure gauge: Analog display

Control: Solenoid valve touch screen automatic operation

Gas purification circulation system (Independent purification caibinet)

Purification column

Function: gas sealed, remove water, remove oxygen

Container material: 304 stainless steel

Purification materials: copper catalyst: 5kg, molecular sieve: 5 kg

Purification capacity: oxygen: 60L, water: 2Kg

Water and oxygen indicators: less than 1ppm

Circulatory system

Working gas: nitrogen, argon

Recycling capacity: integrated fan flow 90m³ / h, install frequency conversion

Regeneration

Operation: PLC automatically controls the regeneration process

Regeneration gas: working gas mixed with hydrogen gas, (hydrogen 5-10%)

Vacuum pump: 8m³ / h, rotary vane pump, with oil mist filter, air vibration control

Valves

Main valve: DN 40 KF, electrically operated angle valve

Control valve: electromagnetic integrated valve (stainless steel integrated valve base, single column for the six valve integration)

Control System

Function

Including self-diagnosis, power-off self-starting characteristics, with pressure control and adaptive function

Automatic control, cycle control, password protection, calculator,

The unit is controlled by Siemens PLC touch screen. The touch screen is locked with a password. It is automatically locked after five minutes of unattended operation. You need to re-enter the password next time you enter the operation interface.

Pressure control: Automatically control the pressure of the cabinet and transition cabin through PLC, cabinet working pressure +/- 10mbar can be set freely, beyond +/- 12mbar system automatic protection

Pedal: Control the box pressure, easy to operate boost and buck

Information memorandum: Automatic recording of system data

Display system

Using Siemens PLC touch screen, display running status, box pressure, system records etc.

The touch screen is locked with a password. It is automatically locked after five minutes of unattended operation. You need to re-enter the password next time you enter the operation interface.

Vacuum system control: Vacuum pump, can be manually or through the PLC to start, flow 8m³/h, vacuum of the transition chamber, and keep the box pressure balance, vacuum pump limit vacuum ≤ 2x10-1pa

Water analyzer

Measuring range: 0 ~ 500ppm

A wide range of applications, especially for lithium manufacturing and metal organic and other users, the water probe can be cleaned by the regeneration process to restore the initial state, can be reused

Oxygen analyzer

Measuring range: 0 ~ 1000ppm

The use of ZrO2 sensor, to avoid the fuel cell life is short, and avoid the problem of it can not be exposed in the air

Organic solvent adsorber: Size: diameter 136mm, height 256mm, filled with 2kg activated carbon, placed inside the box, can quickly replace the material

Datasheet

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts