Temperature range (*1): 5℃ above RT to +70℃
Settable temp. range (*2): -20℃ to +100℃
Temp. control accuracy (*3): ±0.1℃ to ±0.3℃
Stirring method in Bath: Jet flow
Flow: Jet blast
Temperature display/Temperature memory: Display: Digital (Changeable Preset/Current value)Memory: 3
Other functions: Buzzer notification when Preset temp. is reached
Automatic tuning
Minimum Water level: 45 mm from the bottom of the bath
Heater: 800 W
Safe devices/protections: Circuit protector, Fuse, Dry-heating protection with float, Sensor error, Short circuit, High/Low temp. sample protection, Water level alarm, Non-volatile memory error, Automatic tuning error, Alarm setting error
Power failure: Switchable Automatic/Manual recovery
Dimensions inside Bath(W×D×H): 230 × 350 × 140 mm
Main unit dimensions(W×D×H): 280 × 109 × 215 mm (When Thermal insulation tray B-type is used: 280 × 475 × 240 mm)
Weight: Approx. 3 kg (6 kg: Thermal insulation tray B-type is included)
Power supply/consumption: AC100V/8.5A (Need a step-down transformer), 105 Wh at 37℃
Standard accessories: Thermal insulation bath B-type × 1 pc (*5)
Catalog
(*1)Max. temp. may not be reached when the optional Hood is not used or depending on the usage conditions.As it is dangerous due to steam when using high temp. with tap water, it is recommended to use our specified Heat medium for High temp.
(*2)Not equipped with a cooling function. Can be used together with our Chiller/Coolnit, Immersion cooler, etc. at below RT (25℃). Use our specified Antifreeze fluid when preset temp. is below 7℃. The usage life of the components may be shortened when used at the upper limit temp.
(*3)The value under the conditions of RT (25℃), AC100 V/50 Hz, Capacity 8 liters water, Preset temp. 37℃, and No heat load with Thermal insulation tray B-type. For EXN-B, this is an actual measured value (0.01 unit).
(*4)Optional Signal cable CA-671 is required to output. Ask us for the Temperature Input.
(*5)Thermal insulation bath B-type is sold separately.