Defib - Energy Measurement
Load Resistance: 50Ω ± 1% non-inductive
Range: 0 - 199.9 Joules
Accuracy: ± 1% of reading ± 1 Joule
Range (High): 200 - 600 Joules
Resolution: 0.1 Joules
Voltage: 0 - 5000 Volts
Current: 0 - 100 Amps
Maximum Pulse Width: 5μs – 120ms
ECG Simulator
ECG Simulation: Hi-level output
ECG Waveforms – Ventricular Arrhythmias
Premature Ventricular Contraction (PVCI): Intermittent 80 BPM, Amplitude 0.50 - 5.00mV (±2%)
Bigeminy (BIG): 80 BPM, Amplitude 0.50 - 5.00mV (±2%)
Trigeminy (TRIG): 80 BPM, Amplitude 0.50 - 5.00mV (±2%)
Ventricular Flutter (VFLT): 240 BPM, Amplitude 0.50 - 5.00mV (±2%)
Ventricular Fibrillation - Coarse (VFBC): 240 BPM, Amplitude 0.50 - 5.00mV (±2%)
Ventricular Fibrillation - Fine (VFBF): 240 BPM, Amplitude 0.50 - 5.00mV (±2%)
Monomorphic Ventricular Tachycardia (MVT): 210 BPM, Amplitude 0.50 - 5.00mV (±2%)
Right-focal Premature Ventricular Contraction: 80 BPM, Amplitude 0.50 - 5.00mV (±2%)
ECG Waveforms – Atrial Arrhythmias
Sinus Arrhythmia (SAR): 20 - 300 BPM, Amplitude 0.50 - 5.00mV (±2%)
Missing Beat (MB): 20 - 300 BPM, Amplitude 0.50 - 5.00mV (±2%)
Atrial Flutter (AFLT): 300 BPM, Amplitude 0.50 - 5.00mV (±2%)
Atrial Fibrillation (AFB): 20 - 300 BPM, Amplitude 0.50 - 5.00mV (±2%)
Paroxysmal Atrial Tachycardia (PAT): 180 BPM, Amplitude 0.50 - 5.00mV (±2%)
Premature Junctional Contraction (PJC): 20 - 300 BPM, Amplitude 0.50 - 5.00mV (±2%)
ECG Performance Waveforms
Sine (SINE): 0.1 - 300Hz, 1.00 – 10.00mV
Square (SQ): 0.1 - 300Hz, 1.00 – 10.00mV
Triangle (TRI): 0.1 - 300Hz, 1.00 – 10.00mV
Sawtooth (SAW): 0.1 - 300Hz, 1.00 – 10.00mV
Inverse Sawtooth (INVSAW): 0.1 - 300Hz, 1.00 – 10.00mV
Pulse (PULSE): 0.1 - 300Hz, 0.50 – 5.00mV
ECG Waveform Output
Low Level: Low Level
Hi Level: Output Jack
Pacer Input
Fix Load: 50Ω
Accuracy: +/- 4% + 10µJ
Over Voltage Protection: 5000V
Variable Load: 50 to 1600Ω in 50Ω steps
Pulse Rate: 5.0 to 800 ppm
Heart Rate Selection: 20 – 300 bpm
Under & Overdrive: 85% (20 bpm min) and 115% (300 bpm max)
Waveform Selection: NSR, VFibC, VFibF, MVT, AFib, Missing Beat, R-Wave detection
Pulse Current Amplitude: 5.00 – 200mA
Accuracy: +/- (1% rdg +0.02mA)
Current Measurements: Average (RMS), Leading Edge, Trailing Edge, Peak (the highest during the pulse)
Pulse Width: 1.00 - 100ms
Pulse Energy: 1µJ – 2.00J
Pacer Refractory Periods
Refractory Period Test: 15 – 500mS (Paced and sensed)
Accuracy: +/- 1ms
Pacer Interference Test (Immunity)
Heart Rate: 20 - 300 bpm
Frequency: 50 or 60 Hz
Noise Level in mV: 0 - 15.0mV
AED Pulse Mode Waveforms:
Normal Sinus Rhythm (NSR): 20 - 300 BPM, Amplitude 1.00mV(±2%)
Asystole (ASYS):
Ventricular Fibrillation - Coarse (VFBC): 240 BPM, Amplitude 1.00mV(±2%)
Ventricular Fibrillation - Fine (VFBF): 240 BPM, Amplitude 1.00mV(±2%)
Monomorphic Ventricular Tachycardia (MVT): 210 BPM, Amplitude 1.00mV(±2%)
Atrial Fibrillation (AFB): 20 - 300 BPM, Amplitude 1.00mV(±2%)
ECG Waveforms - Sinus:
Normal Sinus Rhythm (NSR): 20 - 300 BPM, Amplitude 0.50 - 5.00mV(±2%)
ST Elevation (STE): 20 - 300 BPM, Amplitude 0.50 - 5.00mV(±2%)
ST Depression (STD): 20 - 300 BPM, Amplitude 0.50 - 5.00mV(±2%)
Myocardial Infarction (MI): 20 - 300 BPM, Amplitude 0.50 - 5.00mV(±2%)
Tall T (TT): 20 - 300 BPM, Amplitude 0.50 - 5.00mV(±2%)
Asystole (ASYS):
ECG Waveforms – Atrial Conduction Arrhythmias:
First Degree AV Block (FAVB): 80 BPM, Amplitude 0.50 - 5.00mV(±2%)
Second Degree AV Block - Mobitz I (SAVB_MI): 80 BPM, Amplitude 0.50 - 5.00mV(±2%)
Second Degree AV Block - Mobitz II (SAVB_MII): 80 BPM, Amplitude 0.50 - 5.00mV(±2%)
Third Degree AV Block (TAVB): 50 BPM, Amplitude 0.50 - 5.00mV(±2%)
ECG Pacer Waveforms:
Synchronous Atrial (AAI): 20 - 300 BPM, Pulse amplitude 0.50 – 5.00mV, Pulse width 0.1 – 2.0ms
Asynchronous Atrial (AOO): 20 - 300 BPM, Pulse amplitude 0.50 – 5.00mV, Pulse width 0.1 – 2.0ms
Pacer (PCR): 20 - 300 BPM, Pulse amplitude 0.50 – 5.00mV, Pulse width 0.1 – 2.0ms
Ventricular Pacer (VVI): 20 - 300 BPM, Pulse amplitude 0.50 – 5.00mV, Pulse width 0.1 – 2.0ms
Atrial & Ventricular Pacer (DDD): 20 - 300 BPM, Pulse amplitude 0.50 – 5.00mV, Pulse width 0.1 – 2.0ms
R-Wave Detection (RWD): 20 - 300 BPM, Pulse amplitude 0.50 – 5.00mV
ECG Noise Selection:
Amplitude: Default Spec Value 0 – 10.00mV
Frequency: 50 - 60Hz
ECG Accuracy:
Rate: Default Spec Value ± 1%
Amplitude: ± 2% (LA-LL), ± 10% (Paddles)
Lead II: 1 – 10 mV (in steps of 0.5 mV)
Other leads are proportional to Lead II by the following percentages:
Lead I: 60%
Lead II: 100%
Lead III: 40%
V1: 63% [Reference LA]
V2: 71% [Reference LA]
V3: 68% [Reference LA]
V4: 80% [Reference LA]
V5: 55% [Reference LA]
V6: 49% [Reference LA]
Pacer Manufacturer Algorithms: CU Medical, GE, HP, Laerdal, Mindray, Philips, PhysioControl, Schiller, WelchAllyn, Zoll
Pacer Sensitivity Test:
Waveform, R Wave: Polarity Normal and Reversed, selectable
Dynamic Sensitivity: 0.05mV to 5.00mV in 50µV steps
General Specifications:
Dimensions: 220mm x 150mm x 90mm (L x W x H)
Weight: 1.5kg
Operation: 9.6V/2400mAh Nickel Metal Hydride battery pack
Battery charge time: 2.5 hours
Battery capacity (fully charged): 12 hours
Mains supply: 110/230V AC; 48 to 66Hz, 35VA power supply
Environmental protection: IP 40
Communication: USB
Display: LCD colour graphic display, ¼” VGA
Memory: 100 test results including graphs
Impact rating: 5J