Size & Weight: 403 × 213 × 433 mm, 11 kg (including power supply)
Power Source: External power adapter: 110–240 VAC – 24 VDC – 220 W; 1 battery
Environmental Operating Conditions: 0°C < T < 50°C; 0% < RH < 95%
Detector Body Maximum Temperature: 185°C
Sample Inlet Maximum Temperature: 185°C
Aspirated Sample Flow: ~800 mL/min at the inlet fitting
Sampling Area Pressure: atm ± 100 mbar
Measuring Range: 0–5,000 mgC/Nm³
Certified Ranges: 0–15 mgC/Nm³; 0–500 mgC/Nm³; 0–1,000 mgC/Nm³; 0–2,000 mgC/Nm³
Accuracy: The greater of 1% of F.S. or 0.4 mgC/Nm³ (0.25 ppm as propane)
Limit of Detection: <0.1 mgC/Nm³
Response Time: 30 sec (with 3 m line)
Method Management & Data Storage: Upload/download methods via user interface; internal flash memory
Interfaces: USB (data download); Wi-Fi (for connection from external device for data visualization and remote control); ModBUS TCP/IP
Flame Management: Electronic control with diagnostics and visual alarm for flame-off and automatic restart system
Hydrogen Storage Cartridge: Metal hydride technology (~20 h continuous operation); rechargeable via external pressurized hydrogen source; no transport restrictions by car/plane (UN 3479)
Zero Air: Non-refillable 1 L cans: technical air @12 bar / 20°C (~10 h continuous operation); quick connectors for external Zero Air cylinder; shipping info: UN 1956, Hazard Class 2.2 (non-flammable, non-toxic gas)
Span Gas: Non-refillable 1 L cans: propane in air @12 bar / 20°C (total carbon equivalent concentration: 16/40/120/320 mgC/Nm³); other configurations on request; quick connectors for external Span Gas cylinder; shipping info: UN 1956, Hazard Class 2.2 (non-flammable, non-toxic gas)