Guaranteed Spec Range:
CO2: 300-500ppm
CO: 0-1ppm
Ch4: 1-3ppm
H2O: 0-7% V
Measurement Technique: Cavity Ring-Down Spectroscopy (CRDS)
Measurement Cell Temperature Control: ±0.005°C
Measurement Cell Pressure Control: ±0.0002 atm
Sample Temperature: -10 to 45°C
Sample Pressure: 300 to 1000 Torr (40 to 133 kPa)
Sample Flow Rate: <0.4 slm at 760 Torr, no filtration required
Sample Humidity: <99% RH non-condensing at 40°C, no drying required
Ambient Temperature Range: 10 to 35°C (operating); -10 to 50°C (storage)
Ambient Humidity: <85% RH, non-condensing
Accessories: Included: external pump, keyboard, mouse; Optional: LCD monitor
Outputs: RS-232, Ethernet, USB, analog (optional) 0–10 V
Fittings: 1/4” Swagelok®
Dimensions: Analyzer: 17” W × 7” H × 17.5” D (43.2 × 17.8 × 44.5 cm), not including 0.5” feet; External pump: 7.5” W × 4” H × 11” D (19 × 10.2 × 28 cm)
Installation: Benchtop (standard) or 19” rack mount chassis (optional)
Weight: 59.3 lbs (26.9 kg), includes external pump
Power Requirements: 100–240 VAC; 47–63 Hz (auto-sensing); <375 W at start-up (total); steady-state operation: 120 W (analyzer), 150 W (pump)