Accuracy: ± 1.5 %
Compliance: CE
Calibration: NIST Certificate
Display Type: LCD
Accuracy Note: ± 1.5% of full scale, including linearity over 15 to 25°C and 5 to 60 psia (0.35 to 4.2 kg/cm2)
Output Detail: Linear 0 to 5 Vdc: 1000 Ω minimum load 4 to 20 mA: 50 to 250 Ω loop resistance
Output Signal: Current, Voltage
Repeatability: ± 0.5%
Response Time: 800 ms
Supply Current: Up to 200 mA
Supply Voltage: 12 Vdc
Turndown Ratio: 100:1
Electrical Output: 4 to 20 mA, 0 to 5 Vdc
Materials, Wetted: 316 Stainless Steel, Anodized Aluminum, Brass, FKM
Pressure Drop, Max: 252 inH₂O
Response Time Note: 800 msec time constant; 2 seconds (typical) to within ± 2% of set flow rate over 25 to 100% of full scale
Media Compatibility: Air, Gases, Nitrogen, Oxygen
Pressure Coefficient: 0.01% FS/psi
Leak Rate Under Vacuum: less than 10^⁻9 cc He/s
Operating Pressure, Max: 1000 psi
Operating Pressure, Min: 0 psi
Process Connection Size: 1/4 in
Process Connection Type: Compression Fitting
Temperature Coefficient: 0.15% FS/°C
Volumetric Flow Rate, Max: 50 SLM
Volumetric Flow Rate, Min: 10 SLM
Operating Temperature, Max: 50 °C
Operating Temperature, Min: 0 °C
Material Construction Option: Aluminum and Brass Body