Weight: 900 g
Stability: ± 0.1% FS typical
Compliance: CE
Weight Note: 900 g (32 oz) Typical depending upon configuration
Battery life: 9 mo
Cable Length: 2.7 m
Display Size: 115 mm (4.5 in)
Display Type: LCD
Accuracy Note: ± 0.08% BSL (0.15% max in 10V/m RF field) (compound gauge calibrated in positive direction only)
Output Signal: Current, Voltage
Pressure Type: Compound Gauge
Pressure, Max: 70 mBar
Pressure, Min: -70 mBar
Wireless Band: None
Accuracy Class: < 0.1%
Burst Pressure: 1015 psi Maximum
Data Interface: USB
Proof Pressure: 10x Full Scale
Supply Voltage: 3.6 Vdc
Pressure Cycles: 1 million minimum
Thermal Effects: Zero Shift: Ranges >350 mb: ± 0.3% spanRanges ≤350 mb: ± 0.5% spanSpan Shift: Ranges >350 mb: ± 0.3% spanRanges ≤350 mb psi: ± 0.5% span
Number of Digits: 4
Display Type Note: 4-digit LCD with user selectable backlighting, (9999 count) 25 mm (1') character height.
Electrical Output: 4 to 20 mA, 0 to 5 Vdc, 0 to 10 Vdc
Enclosure Ratings: IP65, NEMA Type 4X, Weatherproof
Materials, Wetted: 316 Stainless Steel
Pressure Accuracy: ± 0.08 %
Software Settings: Software Settings (via USB) Update Rate: 4 options, 10, 5, 2, 1 times per second Units: psi, inHg, inH2O, bar, mbar, hPa units; must be specified at time of order Lock: Allows the lock-out of front panel functions Alarms: User selectable “High” and “Low” alarm limits (open collector) Analog Output: User Selectable 0 to 5 Vdc, 0 to 10 Vdc, or 4 to 20 mA (requires external 24 Vdc power) Backlight: On/off, 10 sec, 30 sec, 1 min, 5 min Wireless Transmitter (Optional): Channel number, transmission rate, alarms, sensor offset, chart recording, data logging Overrange Indication: “_ _ _ _”
Supply Power Type: Battery
Enclosure Material: 316 Stainless Steel
Number of Batteries: 1
Sample/Display Rate: 0.25 to 10 sec
Settable Parameters: Backlight, Min, Max, Zero, Clear
Thermal Span Effect: ± 0.3%
Thermal Zero Effect: ± 0.3%
Battery Size and Type: 3.6V Li-ion
Sensor Dimension Note: 72 to 89 mm L (2.8 x 3.5") below display
Electrical Output Note: User selectable 0 to 5 Vdc, 0 to 10 Vdc or 4 to 20 mA (external 24 Vdc power supply required) Note: Compound gauge ranges have a unidirectional analog output equal to their full span; i.e. -350 to 350 mb = 0 to 10 Vdc.
Software Compatibility: Requires Windows® 7 (32 bit); XP or Vista
Process Connection Size: 1/4 in
Process Connection Type: BSPP Male
Process Temperature, Max: 66 °C
Process Temperature, Min: -18 °C
Sample Display Rate Note: User selectable from 0.25 to 10 sec. (preset to 1/sec)
Wireless Option Specifications: Transmit Sample Rate: User programmable from 1 sample/2 min to 1 sample/2 sec Radio Frequency (RF) Transceiver Carrier: ISM 2.4 GHz RF Output: 10 dBm (10 mW) Range of RF Link: Outdoors (Line of Sight): Up to 120 m (400') Indoors/Urban: Up to 40 m (130') Included Software: Requires Windows® 7 (32 bit); XP or Vista Internal Battery: Two 3.6V lithium 8.4 Ah capacity (C-cell) included, (replacement battery BATT-C-3.6V) Data Transmission to Host: Pressure reading, ambient temperature reading, RF transmit strength and battery level.