Voltage
- Output range/channel
+ Charge: 0V~5V
+ Discharge: 0V~5V
- Min discharge voltage: 0V
- Accuracy: ±0.02% of range
- Stability: 0.005% of FS
Current
- Output range/channel: Range1: 0.1μA-150μA; Range2: 150μA-5mA; Range3: 5mA-150mA; Range4: 150mA-5A
- Accuracy: ±0.02%of FS
Range1: ±30nA; Range2: ±1μA; Range3: ±30μA; Range4:±1mA
- Stability: 0.005% of FS
Power
- Output power/channel: 25W
- Stability: ± 0.02% of FS
Temperature
- Temperature probe: K-type thermocouple
- Measurement range: -40 ~ +110℃
- Measurement accuracy: ±1℃
- Resolution: 0.1℃
- Date Record Frequency: 1Hz
Time
- Current response time: <= 100μS (10% to 90% or 90% to 10%);
- Working step time: ≤(365*24)h/step Time format-00:00:00.000(h、m、s、ms)
Data Record
- Data record conditions:
+ Time Δt: >=1ms
+ Voltage ΔU: >= 1mV
+ Current ΔI: >= 100nA
- Frequency: 1000Hz (Continuous charge and discharge mode) / Pulse bypulserecording (GSM pulse mode);
Charge modes
- Charge modes: CC, CV, CCCV, CP, CR
- Cut-off condition: Voltage、Current、-△t、Capacity、Energy、Power
Discharge
- Discharge modes: CC、CV、CR、CP、Pulse
- Cut-off condition: Voltage、Current、-△t、Capacity
Pulse Mode
- Charge: CC Mode
- Discharge: CC Mode
- Minimum pulse width: 400μs
- Pulse number: A single pulse step supports 16 different pulse segments
- Cut-off condition Voltage、-△t
DCIR Test: Support DCIR step