Input Functions:
- DC Voltage (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
200 mV/ -20 to 220 mV/ 10 mV/ 0.02 + 0.02/ Impedance approximately 100 MΩ
5 V/ -0.5 to 5.5 V/ 0.1 mV/ 0.02 + 0.01/ Impedance approximately 1 MΩ
50V/ -5 to 55 V/ 1 mV/ 0.03 + 0.01/ Impedance approximately 1 MΩ
- DC mA (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
50 mA/ -4 to 55 mA/ 1 mA/ 0.02 + 0.01/ Impedance : 5 Ω
- Frequency (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
+ 500 Hz/ 3 to 500 Hz/ 0.01 Hz/ ±2 digit/ Impedance: 100 kΩ Min; Sensitivity: 3 Vp-p min; Duty Cycle: 50%
+ 5 KHz/ 3 Hz to 5 KHz/ 0.1 Hz/ ±2 digit/ Impedance: 100 kΩ Min; Sensitivity: 3 Vp-p min; Duty Cycle: 50%
+ 50 KHz/ 3 Hz to 50 KHz/ 1 Hz/ ±2 digit/ Impedance: 100 kΩ Min; Sensitivity: 3 Vp-p min; Duty Cycle: 50%
Continuity (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
<250 Ω/ Close / open/ —/ —/ Open Circuit Voltage : about 2.5 V
Output Funtions:
- DC Voltage (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
+ 1000 mV/ -100 to 1100 mV/ 10 mV/ 0.02 + 0.01/ Maximum output : 2 mA
+ 10 V/ -1 to 11 V/ 0.1 mV/ 0.02 + 0.01/ Maximum output : 5 mA
- DC Current (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
20 mA/ 0 to 22 mA/ 1 mA/ 0.02 + 0.02/ Simulator Transmitter: 5 to 28 V Power Supply outside; 1 KΩ at 20 mA
- Frequency (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
+ 100 Hz/1 to 110 Hz/ 0.01 Hz/ ±2 count/ Output voltage : 1 to 11 Vp-p (zero base waveform); Amplitude Accuracy : ±10% + 0.5 V; Maximum load : >100 KΩ; Duty Cycle : 50%
+ 1 KHz/0.1 to 1.1 KHz/ 1 Hz/ ±2 count/ Output voltage : 1 to 11 Vp-p (zero base waveform); Amplitude Accuracy : ±10% + 0.5 V; Maximum load : >100 KΩ; Duty Cycle : 50%
+ 10 KHz/ 1 to 11 KHz/ 0.1 KHz/ ±2 count/ Output voltage : 1 to 11 Vp-p (zero base waveform); Amplitude Accuracy : ±10% + 0.5 V; Maximum load : >100 KΩ; Duty Cycle : 50%
- Pulse (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
+ 100 Hz/ 1 to 100000 cycles/ 1 cycle/ _/ Output voltage : 1 to 11 Vp-p (zero base waveform); Amplitude Accuracy : ±10% + 0.5 V; Maximum load : >100 KΩ; Duty Cycle : 50%
+ 1 KHz/ 1 to 100000 cycles/ 1 cycle/ _/ Output voltage : 1 to 11 Vp-p (zero base waveform); Amplitude Accuracy : ±10% + 0.5 V; Maximum load : >100 KΩ; Duty Cycle : 50%
+ 10 KHz/ 1 to 100000 cycles/ 1 cycle/ _/ Output voltage : 1 to 11 Vp-p (zero base waveform); Amplitude Accuracy : ±10% + 0.5 V; Maximum load : >100 KΩ; Duty Cycle : 50%
-Loop (Reference/ Ranges/ Resolution/ Accuracy/Remarks):
24 V/ _/ _/ ±10%/ Maximum output current : 25 mA; Short circuit protected
Standard delivery:
Process Meter – Model : Nagman 15+
5V ‘AA’ Alkaline Batteries (4 Nos.)
Operation Manual
Test Leads
Traceable Calibration Certificate
Optionals:
Calibration Certificate from NABL Accredited –as per ISO/IEC 17025:2005 Laboratory