Input
* Voltage (DC) (Range/ Input range/ Resolution/ Accuracy/ Remarks):
100 mV/ -10 to 110 mV/ 10 μV/ ±0.05% measured value ± 30 μV/ Input Resistance 1MΩ
* Thermocouple (Range/ Input range/ Resolution/ Accuracy/ Remarks):
+ R/ -40 to 1760°C/ 1°C/ ±0.05% measured value ±3°C (≤100°C); ±0.05% measured value ±2°C (>100°C)/ Input Resistance 1MΩ
+ S/ -20 to 1760°C/ 1°C/ ±0.05% measured value ±3°C (≤100°C); ±0.05% measured value ±2°C (>100°C)/ Input Resistance 1MΩ
+ B/ 400 to 1800°C/ 1°C/ ±0.05% measured value ±3°C (400 to 600°C); ±0.05% measured value ±2°C (>600°C)/ Input Resistance 1MΩ
+ E/ -200 to 1000°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ
+ K/ -200 to 1370°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ
+ J/ -200 to 1200°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ
+ T/ -200 to 400°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ
+ N/ -200 to 1300°C/ 0.1°C/ ±0.05% measured value ±2°C (≤ -100°C)/ ±0.05% measured value ±1°C (> -100°C)/ Input Resistance 1MΩ
Output
* Voltage (DC) (Range/ Output range/ Resolution/ Accuracy/ Remarks):
100 mV (Low)/ -10 to 110 mV/ 0.01 mV/ ±0.05% set value ±30 μV/ Max. output
current ±2 mA
1V (High)/ -0.10 to 1.10 V / 0.1 mV/ ±0.05% set value ±0.3 mV/ Max. output
current ±2 mA
* Thermocouple (Range/ Output range/ Resolution/ Accuracy/ Remarks):
+ R/ -40 to 1760°C/ 1°C/ ±0.05% set value ±3°C (≤100°C); ±0.05% set value ±2°C (>100°C)/ By using ITS-90 temperature scale
+ S/ -20 to 1760°C/ 1°C/ ±0.05% set value ±3°C (≤100°C); ±0.05% set value ±2°C (>100°C)/ By using ITS-90 temperature scale
+ B/ 400 to 1800°C/ 1°C/ ±0.05% set value ±3°C(400 to 600°C); ±0.05% set value ±2°C (>600°C)/ By using ITS-90 temperature scale
+ E/ -200 to 1000°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale
+ K/ -200 to 1370°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale
+ J/ -200 to 1200°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale
+ T/ -200 to 400°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale
+ N/ -200 to 1300°C/ 0.1°C/ ±0.05% set value ±2°C(≤100°C); ±0.05% set value ±1°C (>100°C)/ By using ITS-90 temperature scale
Standard delivery:
Thermocouple Calibrator – Model : Nagman 02
9V Alkaline Battery
Operation Manual
Test Leads
Traceable Calibration Certificate
Optionals:
Calibration Certificate from NABL Accredited –as per ISO/IEC 17025:2005 Laboratory