Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Weight: 4.5 kg (10 lbs)
Outputs: 4 Analog
Humidity: 10-90% non-condensing
Flow Rate: 40 mL/min – 60 mL/min
Sample pH: 2.5-12
Update Rate: Once per second
Power Supply: 100-240 VAC, 50-60 Hz, 25 W, on power loss all settings are retained without batteries
Analog Output: 22mA alarm, 500 ohm max load, not for use with externally powered circuit For sodium, conditioned pH, temperature four powered 0/4-20 mA,
EnclosureType: Full
Response Time: 90% at 5 minutes
Relay Contacts: 3A resistive freely assignable to setpoints for sodium, pH, temperature other relays used for auto-cal
Operating Range: 0.01 ppb – 100,000 ppb
Sample Pressure: 0.3-7 bar (5-100 psig)
Sample Flow rate: >40 mL/min (>20 mL/min for cation exchange samples), excess to drain
Ratings/Approvals: CE, UL, IP65
Built-in sequencer: No
Dimensions (HxWxD): 851 mm x 400 mm x 165 mm
Operator Interface: 4-line backlit LCD, 5 tactile keys simultaneous display of sodium, conditioned pH, auto-cal status (temperature optional)
Process Connection: Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included Sample inlet: 6 mm or 1/4" OD tube SS compression fitting Drain hose: 19 x 25.4 mm (¾ x 1"), 2 m (6 ft) length included
Sample Temperature: 5-50°C (41-122°F)
Sodium Calibration: Manual 3-point known addition manual 1- or 2-point
Reagent Consumption: 3-4 L per year DIPA, or 9-12 L per year ammonium hydroxide typical more at higher temperatures and for cation exchange samples
Power Supply (AC/DC): 100 V – 240 V
Analog Output Scaling: Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging
Measurement Parameter: Sodium
Analog Output Accuracy: ± 0.05 mA
Electrode Conditioning: Manual
Grab Sample Measurement: Available option
Operating Range (Temp.): 5 °C – 50 °C
Sodium Measurement Range: 0.01-100,000 ppb or equivalent ppm, auto-ranging
Sodium Measurement Accuracy: ± 10% of reading ± 0.1 ppb, typical using DIPA as reagent ± 10% of reading ± 1 ppb, typical using ammonium hydroxide as reagent
Ambient Operating Temperature: 5-50 °C (41-122 °F)