Weight: 27 kg (60 lbs)
Humidity: 10-90% non-condensing
Sample pH: 2.5-12
Calibration: Automatic
Update Rate: Once per second
Model Family: 2300Na
Power Supply: 100-240 VAC, 50-60 Hz, 25 W, on power loss all settings are retained without batteries
Analog Output: 22mA alarm, 500 ohm max load, not for use with externally powered circuit For sodium, conditioned pH, temperature four powered 0/4-20 mA,
EnclosureType: Full
Response Time: 90% at 5 minutes
Relay Contacts: 3A resistive freely assignable to setpoints for sodium, pH, temperature other relays used for auto-cal Two unpowered, SPDT, 250VAC/30VDC
Sample Pressure: 0.3-7 bar (5-100 psig)
Sample Flow rate: >40 mL/min (>20 mL/min for cation exchange samples), excess to drain
Built-in sequencer: No
Dimensions (HxWxD): 35.4"x17.7"x7.5" (900x450x190 mm)
Operator Interface: 4-line backlit LCD, 5 tactile keys simultaneous display of sodium, conditioned pH, auto-cal status (temperature optional)
Process Connection: Sample Inlet: 1/4" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3/4 x 1") 2 m (6 ft) length included
Sample Temperature: 5-50°C (41-122°F)
Sodium Calibration: Automatic, unattended 3-point known addition
Reagent Consumption: Diisopropylamine (DIPA), approximately 1 L per 3 months more at higher temperatures and for cation exchange samples
Analog Output Scaling: Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging
Measurement Parameter: Sodium
Analog Output Accuracy: ±0.05 mA
Electrode Conditioning: Part of auto-cal sequence
Grab Sample Measurement: Included
Sodium Measurement Range: 0.001 - 100,000 ppb or equivalent ppm, autoranging
Sodium Measurement Accuracy: ±10% of reading or ± 0.05 ppb whichever is greater
Ambient Operating Temperature: 10-45°C (50-113°F)