Interior dimension: 1250W x 1450H x 750D mm
Exterior dimension: 1435W x 1913H x 905D mm
Temperature
Setting temperature range without light, with humidity: +15 to 60 °C
Setting temperature range without light, without humidity: 0 to +70 °C
Working temperature range without light, with humidity: +15 (at least 10 below ambient temperature) to +60 °C
Working temperature range without light, without humidity: 0 (at least 20 below ambient temperature) to +70 °C
Setting accuracy temperature: 0.1 °C
Temperature sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Humidity
Setting range humidity: 10 - 80% rh
Humidity: humidity supply with distilled water from external tank by self-priming pump
Humidification: humidification by hot steam generator
Dehumidification: dehumidification by cold trap using the Peltier technology
Setting accuracy humidity: 0.5% rh
Control technology
ControlCOCKPIT: TwinDISPLAY. Adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-colour displays.
Language setting: German, English, Spanish, French, Polish, Czech, Hungarian
Adjustable parameters: temperature (Celsius or Fahrenheit), relative humidity, programme time, time zones, summertime/wintertime
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Function HeatBALANCE: adapting the distribution of the heating performance of the upper and lower heating circuit from -50% to +50%
Function SetpointWAIT: the process time does not start until the set temperature is reached
Calibration: three freely selectable values each, temperature and humidity
Communication
Interface: Ethernet LAN, USB
Documentation: programme stored in case of power failure
Programming: AtmoCONTROL software on a USB stick for programming, managing and transferring programmes via Ethernet interface or USB port
Safety
Temperature control: over- and undertemperature monitor TWW, protection class 3.3 or adjustable temperature limiter TWB, protection class 2, selectable on display
AutoSAFETY: additionally integrated over- and undertemperature protection "ASF", automatically following the setpoint value at a preset tolerance range, alarm in case of over- or undertemperature, heating function is switched off in case of overtemperature, cooling function in case of undertemperature
Autodiagnostic system: integral fault diagnostics for temperature and humidity control
Alarm: visual and acoustic
Peltier: energy-saving Peltier heating-/cooling system integrated in the rear (heat pump principle)
Standard equipment
Standard works calibration certificate: +25°C / 60% rh, +40°C / 75% rh
Door: Stainless steel doors with glass sectors, fully heated inner glass panes integrated in the full-sight glass door with 2-point locking (compression door lock)Door: inner glass door
Internals: 4 stainless steel grid(s), electropolished
Standard accessories: Water tank including connection hose
Electrical data
Voltage: 230 V, 50/60 Hz
Electrical load: approx. 1400 W
Ambient conditions
Ambient temperature: 16 ºC to 40 ºC
Humidity rh: max. 70 %, non-condensing