General

Measurable Materials: Semi-hard materials and Cemented Carbides

Measurable Quantities: Msat, Hc, Jsat, weight-specific saturation moment (σsat), %Co 

Main Cabinet

Accuracy: Gaussmeter: 0.25% reading + 0.1% range; Probe: 0.5% linearity 

Resolution: 0.1 Oe (300 Oe range), 1 Oe (3000 Oe range) 

Hc Measurement: Accuracy: ±2% for Hc > 500 A/m

Ms Measurement: Accuracy: ±2% on readin; Resolution: 10⁻⁴ µWbm (10⁻¹⁰ Tm³, 10⁻¹ emu) 

Sample Size 

With Coil 5800 MC-1: Ø 37 mm x H 19 mm (1.45” dia x 0.75” H) 

With Coil 5800 MC-2: Ø 27 mm x H 13.6 mm (1” dia x 0.6” H) 

Poles Diameter: 100 or 120 mm (4” or 4.7”)

Accuracy: Max H Field 

With LP-100 mm Pole: 9700 Oe (776 kA/m) 

With LP-120 mm Pole: 7750 Oe (620 kA/m) 

Test Time: 1 minute (typical) 

Operating Temperature Range: 10°C to 35°C

Communication Port: RS-232 / USB2 

Main Electrical Cabinet

Power Supply: 220V, 50/60 Hz

Dimensions: 535 x 655 x 550 mm (21 x 26 x 22”)

Weight: 58.5 kg (129.3 lb) 

Fluxmeter

Ranges: (1,2,5,10,20,50,100) x 2000 µWb

Resolution: From 1 µWb (range 1) to 100 µWb (range 100)

Accuracy: ±0.5%

Drift: 10 kΩ x range

Communication: RS232 / USB 