Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Types
+ UCAM-60B-AC AC only
+ UCAM-60B-DC DC only
TEDS:
+ Interface: IEEE 1451.4 Mixed Mode Transducer Interface Class 2
+ Applicable sensor: Should have information written in accord-ance with IEEE template No. 33; cable length should be 30 m or less
Scanning Speed
Notes:
1. High-resolution mode and high-speed mode are selectable for dedicated scanners only.
2. High-resolution or high-speed mode is available only with full bridge system.
File Conversion: CSV conversion
Coefficient Calculation: Multiplication by calibration coefficients, calibration by TEDS, conversion of measured values to physical quantities, scaling and correction.
Self-diagnosis Functions: Checks display, printer, bridge excitation, lead-wire-off, input/output resistance, insulation resistance, mode, etc. Checking of input/outut resistance and mode is availabe only for dedicated scanners.
Automatic Measurement Functions (Trigger Measurement): A relative value (Certain changing quantity) or an absolute value triggers measurement. In addition to the usual trigger function, a variable trigger function is provided with which the trigger value changes at each step during measurement. With this special function, a trigger value and the number of measurement times (Repeat times) under the trigger condition can be registered for each step to perform a series of automatic measurements in the order of steps.
Automatic Measurement Functions (Interval MeASUREMENT: ): Measurement is automatically performed at preset time intervals.
Automatic Measurement Functions (Trigger Interval Measurement): Combination of trigger measurement and interval measurement.
Channels: Max. 30 with dedicated scanners; Max. 1000 with external scanners connected
Measuring Functions:
+ Initial: Initial values are measured and stored in internal memory (Except for temperatures measured by civil engineering transducers with temperature measuring function).
+ Original: Raw values are measured without subtraction of initial values.
+ Measure: Initial values are subtracted from original values (Except for temperatures measured by civil engineering transducers with temperature measuring function).
+ Easy Measure: Auto zero balancing function is activated.
+ Note: The selected function is applied to all channels. Coefficient
Measuring Targetss: Strain gages, strain gage transducers, civil engineering transducers with temperature measuring function, DC voltage-output or DC current-output instruments, potentiometer sensors, thermal sensors (Thermocouples and platinum resistance thermometer bulbs)
Standard Accessories:
+ AC power cable P-18 with 2-pin conversion plug CM-39 (AC-operated version)
+ DC power cable P-76 (DC-operated version)
+ Recording paper UCAM-60A-RP (1 roll for UCAM-60B only)
+ Screwdriver
+ Spare fuse
+ CD-R (Instruction Manual)
Optional: Recording Paper UCAM-60A-RP (10 rolls/pack)
Sacnning Speed: USB-70B (Standard Mode only) : Standard Mode (50 ms/channel) Besides these, the following speeds can be set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s and 10 s
Auto Measuring Functions (Trigger Measurement): Trigger channel : 1 desired channel
Automatic Measurement Functions (Trigger Measurement): Trigger Value : A desired real number of 6 effective figures or less Reference
Auto Measuring Functions (Trigger Measurement):
+ standard Value : Amount of level shift to determine the first trigger value (Selected from the same range as for the trigger values)
+ Repeat Times : 1 to 9999 (0 for infinite times)
+ Measuring Steps : Maximum 15
Auto Measuring Functions (Interval Measurement):
+ Measuring Start Time : Year/month/day and hour: minute: second
+ Interval setting: Days and hours: minutes: seconds (In a range of 00 and 00: 00: 01 to 99 and 12: 59: 59)
+ Measuring Interval : Days and hours: minutes: seconds (In a range of 00 and 00: 00: 01 to 99 and 12: 59: 59)
+ Repeat Times : 1 to 9999 (0 for infinite times)
+ Available Steps : Maximum 15
Auto Measuring Functions (Trigger Interval Measurement):
+ Measuring Times : Max. 9999
+ Interval Time : Available in a range of 1 second to 99 days 23 hours: 59 minutes: 59 seconds
Storage Functions:
+ Built-in Memory : Internal memory, approx. 7 MB
+ Exterior Memory : Flash ATA card (Optional); the capacity depends on the card.
Sacnning Speed:
+ Exclusive Scanning Unit (Standard Mode): : 50 ms/channel. Scanning speeds stated above are standard maximum speeds in respective modes.
+ Besides these, the following speeds can be set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s and 10 s
+ Exclusive Scanning Unit (High-Conversion Mode): : 280 ms/channel (High-resolution mode)
+ Note: Individually selectable for desired channels
+ Exclusive Scanning Unit (High-speed Mode): : 20 ms/channel (High-speed mode), voltage, Current, only Potentiometer sensors