Accuracy and repeatability equivalent to ASTM approved CFR engine test methods(ASTM D2699, D2700, D4737, D613; ISO 4264, 5163, 5164; IP 236, 237)
System Capabilities:
Pre-calibrated for: Octane number RON, MON, (R+M)/2
Optional Calibrations: Diesel cetane index, cetane number, biodiesel percentage
Outputs: Full Color Display, built-in Printer that includes time and date of analysis
Data Transfer: Via USB port for interface with PC
Data Analysis: Includes full calibration software
Results: Traceable to CFR Engine or other fuel lab standards
Calibrations: Stores up to 10 calibration sets
Automatic fuel detection (Gasoline/E-Gas/Diesel)
Automatic temperature compensation
Self-standardizing to ensure accuracy
Data Storage: Stores data for up to 100 samples
Optical Capabilities:
Measurement Mode: Diffuse transmittance
Spetrum Range: 16 filters covering wavelengths from 810 to 1045 nm
Scan Speed: Up to 10 scans per second
Optical Range: 0 to 6 AU
Resolution: 0.00001 AU
Stability: 0.05 Mili-AU
Measurement Time: Variable (Typically under 25 seconds)
Sample Information:
Sample Size: 200mL with 75mm pathlength
Sample Holder: Reusable glass with chemical seal cover
Sample Preparation: None required
Sample Destruction: None
Dimensions: 4.5” x 13.5” x 2.2” in (12cm x 33cm x 5cm)
Net Weight: 3 lbs. (less than 2 kg)
Electrical Requirements: 240V AC 50/60Hz
Included Accessories
Instrument
Case
AA Batteries
5 Rolls of Printer Paper
AC Power Cord
Calibration Software
User’s Manual
Quick Start Guide
Calibration Manual