Measuring Gas: Combustible gas(Methane) in the air
Measuring Principle: Catalytic combustion
Measuring Range: 0 - 100%LEL
Resolution: 1%LEL
Indication accuracy: Within ±10%LEL
Alarm accuracy: Within ±10%LEL
Alarm concentration: Factory default settings
AL1 : 20%LEL
AL2 : 0%LEL(alarm clear)
Response Time: Within 25 seconds at 90% response from sampling gas inlet at 20°C
Measuring Gas: Oxygen in the air
Measuring Principle: Galvanic cell
Measuring Range: 0.0 - 50.0%
Resolution: 0.1%
Indication accuracy: 0.0 - 25.0 % : ±0.5%; Above 25.1% : ±3.0%
Alarm accuracy: ±0.5vol% against alarm setting value 18.0vol%
Alarm concentration: Factory default settings
AL1 : 18.0%
AL2 : 0.0%(alarm clear)
Response Time: Within 15 seconds at 90% response from sampling gas inlet at 20°C
General
Sampling method: Suction type Approx.0.6L/min (at room temperature with 5m gas sampling hose)
Display: Digital liquid crystal
Alarm Method: LED lamp, LCD display and buzzer
Pressure range: 80 - 110 kPa
Power Supply: 3 x size AA Alkaline dry batteries(LR6)
Continuous Operating time: Approx. 8 hours, no alarm
Explosion proof: ExiadII CT4X
Operating Temperature Range: -10-40°C; below 95%RH (non-condensing)
Main body dimensions: 78(W) x 200(H) x 50(D)mm
Main body weight: Approx. 550g (including batteries)
Standard accessories: Gas sampling hose with float type gas collector(5m, polyurethane), leather case, carrying case
Optional accessories: Gas sampling hose with float type gas collector(10m, 20m, 30m, polyurethane)