Measuring Gas: Carbon monoxide
Measuring Principle: Chronoamperometry
Measuring Range: 0 - 500ppm
Resolution: 1ppm
Detectable accuracy: Below 100ppm: ±10ppm; Above 101ppm: ±10% of indicated value
Alarm accuracy: Below 100ppm: ±10ppm; Above 101ppm: ±10% of indicated value
Alarm concentration: Factory default settings; AL1: 50ppm; AL2: 100ppm
Response Time: Within 25 seconds at 90% response from sampling gas inlet at 20°C
Measuring Gas: Oxygen
Measuring Principle: Galvanic cell
Measuring Range: 0.0 - 50.0vol%
Resolution: 0.1vol%
Detectable accuracy: 0.0 - 25.0vol% : ±0.5vol%; Above 25.1vol% : ±3.0vol%
Alarm accuracy: ±0.5vol% against alarm setting value 18.0vol%
Alarm concentration: Factory default settings; AL1 : 18.0%
Response Time: Within 25 seconds at 90% response from sampling gas inlet at 20°C
General
Sampling method: Suction type Approx.0.6L/min (at room temperature with 5m gas sampling hose)
Display: Digital liquid crystal
Alarm Method: LED lamp, LCD display and buzzer
Operating Temperature Range: -10-40°C; 30-85%RH (non-condensing)
Pressure range: 80 - 110 kPa
Power Supply: 3 x size AA Alkaline dry batteries(LR6)
Continuous Operating time: Approx. 16 hours, no alarm
Explosion proof: ExiaII CT4X
Main body dimensions: 78(W) x 200(H) x 50(D)mm
Main body weight: Approx. 550g (including batteries)
Standard accessories: Gas sampling hose with float type gas collector(5m, fluorine-contained), leather case, carrying case
Optional accessories: Gas sampling hose with float type gas collector(10m, 20m, 30m, fluorine-contained), assembly for O2 sensor replacement, assembly for CO sensor replacement