Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

DC Ratings:
Voltage 80 V
Current 40 A
Power 2 kW
Programming Accuracy:
Voltage 0.03% + 24 mV
Current 0.1% + 40 mA
Resistance 3 mΩ
Power 0.5% + 10 W
Readback Accuracy
Voltage 0.03% + 24 mV
Current 0.1% + 60 mA
Power 0.5% + 10 W.
Load Regulation
Voltage ≤ 16 mV
Current ≤ 40 mA.
Note: Percent of value + offset; at 23 °C ± 5 °C after a 30-minute warm-up; measurement NPLC=1; valid for 1 year.
Constant Voltage Mode Programming and Readback
Range 0 to 80 V
Resolution 1 mV
Temperature coefficient ≤ 30 ppm/°C.
Constant Current Mode Programming and Readback
Range 0 to 40 A
Resolution 10 mA
Temperature coefficient ≤ 50 ppm/°C.
Constant Resistance Mode Programming:
Load resistance range 0.03 to 2,400 Ω
Resolution 1 mΩ.
Constant Power Mode Programming and Readback
Range 0 to 2 kW
Resolution 1 W.
Maximum Slew Rate
Rise/Fall slope 10 A/ms.
Line Regulation
Voltage ≤ 16 mV
Current ≤ 24 mA.
Protection Capability
Overvoltage 82 V
Overcurrent 42 A
Overpower 2,040 W.
*For detailed specifications, please refer to the Datasheet.