Internal Dimension WxHxD (mm): 800x800x800
Dimension External Dimension WxHxD (mm): 1200x1880x2050
Volume (L): 512
Pressure:
Pressure range: Normal pressure~0.5Kpa (500pa)
Pressure error: Ordinary pressure~40kpa±2kpa;40kPa~4kPa ≤±5kPa; 4kPA~1kPa ≤±0.1kpa
Depressurization time: From site level to 15250m in 10minutes; From site level to 30500m in 30minutes; From site level to 22860m in 20minutes;
Rapid decompression time: 75.2kPa~18.8kPa≤15s (Optional)
Explosive decompression time: 75.2kPa~18.8kPa ≤0.1s (Optional)
Temperature:
Temperature range:
+ Low temperature range: -40°C
+ High temperature range: 150°C(180°C is Optional)
Temperature fluctuation : ±0.5°C
Temperature Rising/ Falling Speed Rate:
+ Temperature rising approx:0.1~2.0°C/min
+ Temperature falling approx:0.3~1.0°C/min; (3°C 5°C 10°C 15°C 20°C 25°C/min with Normal pressure is optional for Whole courseaverageor Linear average);
Temperature uniformity: ≤±2.0°C(Normal pressure, no load, constant state)
Humidity:
Humidity range: 20% RH~98% RH (work under atmospheric humidity 10% R.H~98% RH (Optional)
Humidity fluctuation : ±2.0% RH
Humidity uniformity: ≤±3.0% RH (humidity >75% RH); ≤±5.0%RH(humidity≤75% RH)
Others:
Power: AC380V 3phase 5lines 50/60Hz
Aprroximate weight(kg): 1000