Parameter: DO
Principle: Phosphorescence
Range: Air saturation 0 to 200%
Resolution: 0.01 to 0.04%
Accuracy: Non-linearity ±2% of full scale (at 1 atm, 25 °C)
Parameter: Temperature
Principle: Thermistor
Range:-3 to 45 °C
Resolution: 0.001 °C
Accuracy: ±0.02 °C (ARO-CAV: 3 to 31 °C) (Others: 0 to 35 °C)
Signal output: CAR: RS-232C, CAD: RS-485
Communication: Handshake
AD converter: 16 bit digital conversion
Power: 12 to 24 VDC
Current drain: 35 mA
Material: Titanium (grade 2)
Dimensions: Φ54 × 185 mm (w/o connector)
Weight: Approx. 0.5 kg in air, 0.3 kg in water
Pressure rating: 1000 m depth equivalent
Connector: Fixed type (standard) or detachable type