For full functionality of this site it is necessary to enable JavaScript.
EMIN.CO.TH
0
Product image

HV HIPOT GDFJ-VI Portable Dissolved Gas Analyzer

ModelGDFJ-VI
OriginChina
Warranty12 months
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

FID (flame ionization detector)

Detection limit: Mt≤3×10-12g/s (Hexadecane/isooctane)

Baseline noise: ≤5×10-14A

Baseline shift: ≤1×10-13A/30min

Linear range: ≥ 106

TCD (thermal conductivity detector)

Sensitivity: S≥3500mV•ml/mg (conventional); 5000mV•ml/mg (highly sensitive)

Baseline noise: ≤10μV

Baseline shift: ≤30μV/30min

Linear range: ≥104

Temperature control index

Temperature control part: Range

Column box: 4~ 450℃ above room temperature, accuracy ±0.1℃

Packed column injector: 4~ 450℃ above room temperature, accuracy ±0.1℃

FID Detector: 4~ 450℃ above room temperature, accuracy ±0.1℃

TCD Detector: 4~ 450℃ above room temperature, accuracy ±0.1℃

Ni nickel catalyst reformer: 4~ 450℃ above room temperature, accuracy ±0.1℃

Other parameters

Dimension: 420*280*300mm

Weight: <15kg

Power supply: AC220V±10%, 50Hz


Accessories

Name Specification Qty.

Portable DGA: GDFJ-VI x 1 set

Chromatography workstation: Insulating oil special edition x 1 set

Miniature Full Automatic Oscillator: x 1 set

High Purity Nitrogen: 2L, 99.999%, Aluminum alloy bottled x 2 bottles

High Purity Hydrogen: 2L, 99.999%, Aluminum alloy bottled x 1 bottle

Standard Gas: 2L, 7 Components, Aluminum alloy bottled x 1 bottle

Datasheet

Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.

Quick Support

Direct access to our certified experts