Rated current : 500 A AC/DC
Frequency bandwidth : DC to 200 kHz
Diameter of measurable conductors : Max. φ 50 mm (1.97 in.)
Max. allowable input : ±1500 Apeak
(Within 20 ms in an environment of 40°C/104°F or less)
Output voltage : 4 mV/A
Output resistance : 50 Ω ± 10 Ω
Accuracy (amplitude) ±(% of reading + % of full scale) DC: 0.2% + 0.02%
DC < f ≤ 100 Hz: 0.2% + 0.01%
Linearity : ±20 ppm Typical
Common-Mode Voltage Rejection Ratio
(CMRR) : DC to 1 kHz:150 dB or greater
1 kHz to 10kHz:130 dB or greater
10 kHz to 100 kHz:100 dB or greater
(effect on output voltage and common mode voltage)+F56
Automatic phase correction : Automatically performs phase correction when connected to PW8001
Operating temperature, humidity : -40°C to 85°C (-40°F to 185°F), 80% RH or less
Standards : Safety IEC 61010-2-032:2012/EN 61010-2-032:2012 Type D
EMC IEC 61326-1:2012/EN 61326-1:2013
Withstand voltage : AC 4,260 V
Rated power supply voltage : ±11 V to ±15 V
Max. rated power : 7 VA or less (When measuring 500 A/55 Hz with ±12 V power supply)
Dimensions and mass : Approx. 238 mm W × 116 mm H × 35 mm D
(approx. 9.37 in. W × 4.57 in. H × 1.38 in. D), 860 g (30.3 oz.)
Cable length : Approx. 3 m (9.84 ft.)
Included accessories : ・Markup bands(×6)
・Carrying case
・Instruction manual