For full functionality of this site it is necessary to enable JavaScript.
EMIN.CO.TH
0
Product image

HACH HQ4100 Portable Multi-Meter

ModelHQ4100
P/NLEV015.53.4100A
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Conductivity measurement

Temperature correction: None; Linear; NaCl Non-Linear Natural Water.

Technical Attributes

Accessories Included: USB charger, power cable, and clips

Automatic Buffer Recognition: Yes

Backlight: Yes

Barometric Pressure Measurement: Automatic compensation of DO when using an LDO or LBOD probe

Calibration:

Demal (1D/ 0.1D/ 0.01D); Molar (0.1M/ 0.01M/0.001M); NaCl (0.05%; 25µS/cm; 1000µS/cm; 18mS/cm); Standard sea water; User defined

Calibration Intervals/Alerts/Reminder: Off, selectable from 2 hours to 7 days

Conductivity Measurement Range: Yes

Custom Calibration Standards: Yes

Data Export: USB connection to PC or USB storage device (limited to the storage device capacity).

Data Memory: 100,000 data points

Data Storage: Automatic in Press to Read Mode and Interval Mode. Manual in Continuous Read Mode.

Dimensions (H x W x D): 63 mm x 97 mm x 220 mm

Display: Up to 3 parameters at a time, dependent on HQ model

Display Type: 640x480 Color TFT

DO Measurement Range: 0.1 - 20.0 mg/L (ppm), 1 - 200% saturation

DO Resolution: 0.1

Enclosure Rating: IP67 (with battery compartment installed)

GLP Features: Date; Time; Sample ID; Operator ID, Calibration

Inputs: 1

Instrument: Portable

ISE Direct Measurement Range: Yes

ISE Electrode Calibration: Yes

Kit: Meter only

Languages: English, Chinese - Simplified, Spanish, French, German, Italian, Japanese, Turkish, Dutch, Portuguese (PT & BR), Korean, Danish, Swedish, Polish, Norwegian, Hungarian, Greek, Finnish, Czech, Romanian, Croatian, Bulgarian, Slovak, Lithuanian, Estonian, Slovenian, Russian

Lock Function: Continuous / Auto-stabilization ("press to read") / At Interval

Measurement Method: Probe specific programmed method settings

Model: HQ4100 Multi/ISE/1 Channel

mV Measurement at Stable Reading: Yes

mV Measurement Range: Yes

mV Resolution: 0.1 mV

Operating Error Messages: Clear text error messages displayed

Operating Humidity: 90% (non-condensing)

Operating Interface: Soft Touch Keypad

Operating Temperature: 0 - 60 °C

ORP Electrode Calibration: Predefined ORP standards (including Zobell's solution)

Parameter: pH, Conductivity, TDS, Salinity, Dissolved Oxygen (DO), ORP, ISE

pH Buffer Sets:

Colour-coded: 4.01, 7.00, 10.01 pH;

IUPAC: 1.679, 4.005, 7.000, 10.012, 12.45

DIN: 1.09, 4.65, 9.23

User-defined custom buffer sets

pH Electrode calibration:

1 - 3 Calibration points

Calibration summary data logged and displayed

pH Measurement Range: 0 - 14 pH

pH Resolution: Selectable: 0.001/0.01/0.1 pH

Power Supply: Rechargeable lithium-ion battery 18650 (internal)

Class II, USB power adapter: 100 - 240 VAC, 50/60 Hz input; 5 VDC at 2 A USB power adapter output (external)

Probe Type: Intellical Standard Laboratory or Rugged Field

Probes included: None

Protection Class: IEC Class III (SELV (Separated/Safety Extra-Low Voltage) powered); external

Salinity Measurement Range: 0 - 42 (ppt) (‰)

Salinity Resolution: 0.01 (ppt) (‰)

Sensor A: None

Sensor B: None

Sensor C: None

Storage Conditions: -20 - 60 °C, max. 90% relative humidity (non-condensing)

TDS Measurement Range: 0.00 mg/L - 50.0 g/L NaCl

TDS Resolution: 0.01 mg/L - 0.1 g/L upon measuring range

Temperature Compensation: Automatic Temperature compensation for pH

Temperature Measurement: °C or °F

Temperature Resolution: 0.1 °C

Type of measurement: Field Measurements

Weight: 530 g (Meter only)

Included: HQ4100 Portable pH/ORP/EC/TDS/DO/ISE Multi-Meter, 1 channel, includes manuals, Li-Ion batteries, USB charger with cable for USA.


Stay Updated with Offers

Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.

By subscribing, you agree to our Terms of Service and Privacy Policy.