Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Manufacturer: Graphtec - Nhật
Model: GL900-8
GL900 main unit specifications
|
Item |
Description |
||
|
GL900-8 |
GL900-4 |
||
|
No. of analog input ch. |
8 ch |
4 ch |
|
|
External input/output |
Trigger input (1 channel), Logic input (4 channels) or Pulse input (4 channels), Alarm output (4 channels) |
||
|
Sampling interval |
10μs to 1 min |
||
|
TIME/DIV |
10 ms/DIV to 24 hour/DIV |
||
|
Timer functions |
Date and time, daily cycle, hourly cycle |
||
|
Trigger functions |
Type |
Start: Data capture starts when a trigger is activated; Stop: Data capture stops when a trigger is activated |
|
|
Condition |
Start: Off, Input signal level (analog, logic/pulse), External*1, Scheduled time |
||
|
Stop: Off, Input signal level (analog, logic/pulse), External *1, Scheduled time |
|||
|
Combination |
Input signal level: Level OR, Level AND, Edge OR, Edge AND |
||
|
Mode |
H (Rising), L (Falling), Window In*2, Window Out*2 |
||
|
Alarm setting functions |
Rising, Falling, Window In*2, Window Out*2 |
||
|
Alarm output *1 |
Number of channels: 4, Open collector output (5V, 10 kΩ pull-up resistance) |
||
|
Pulse input*1, *3 |
RPM mode |
50 to 20 M RPM/F.S. (in steps of 1, 2, or 5) |
|
|
Count mode |
50 to 20 M C/F.S. (in steps of 1, 2, or 5) |
||
|
Inst. Mode |
50 to 20 M C/F.S. (in steps of 1, 2, or 5) |
||
|
Calculation functions |
Statistical calculations *4: Average, Peak, Maximum, Minimum, RMS (2 calculations can be set simultaneously) |
||
|
Other functions |
Search function, annotation input function |
||
|
PC inteface |
Ethernet (10BASE-T/100BASE-TX), USB (High Speed supported) provided as standard |
||
|
Ethernet functions |
Web server function, FTP server function, NTP client function |
||
|
USB function |
USB drive mode (File transfer and deletion from internal GL900 memory) |
||
|
Memory device |
Internal |
One million data points / Internal flash memory:Approx. 256 MB |
|
|
External |
USB memory slot (High speed supported) *5 |
||
|
Display screens |
Waveforms + digital values, enlarged waveforms, digital values + calculation results, X-Y |
||
|
Display unit |
5.7-inch TFT color LCD |
||
|
Operating environment |
0 to 40℃, 5 to 85% R.H. (15 to 35℃ when using batteries) |
||
|
Withstand voltage |
Between each input channel and GND: 1000 V p-p for one minute, between input terminals: 1000 Vp-p for one minute |
||
|
Power supply |
AC adapter |
100 to 240 VAC, 50 to 60 Hz |
|
|
DC input |
8.5 to 24 VDC |
||
|
Battery pack *6 |
Option |
||
|
Power consumption |
42VA (when operating and charging battery with AC power) |
40VA (when operating and charging battery with AC power) |
|
|
External dimensions |
232 x 150 x 80 mm (W x H x D), approx. |
||
|
Weight (approx.) |
1.1 kg (excluding AC adapter and battery) |
1.0 kg (excluding AC adapter and battery) |
|
|
Vibration-tested conditions |
Equivalent to automobile parts Type 1 Category A classification |
||
Terminal block specifications
|
Item |
Description |
|||
|
GL900-8 |
GL900-4 |
|||
|
Number of input channels |
Fixed to 8 channels |
Fixed to 4 channels |
||
|
Input terminal type |
Voltage |
BNC connector |
||
|
Temperature |
M3 screw type terminal board *7 |
|||
|
Input method |
All channels isolated Imbalanced input Simultaneous sampling of all channels |
|||
|
Measurement ranges |
Voltage |
20, 50, 100, 200, 500 mV; 1, 2, 5, 10, 20, 50, 100, 200, 500 V F.S., 1-5 V F.S. |
||
|
Temperature |
Thermocouples : K, J, E, T, R, S, B, N, W (WRe5-26) |
|||
|
Humidity |
0 to 100% (voltage 0 V to 1 V scaling conversion) * with B-530 (option) |
|||
|
Input filter |
Off, Line, 5 Hz, 50 Hz, 500 Hz |
|||
|
Measurement accuracy *8 |
Voltage |
±0.25% of F.S. |
||
|
Thermocouple |
Type |
Measurement temperature range |
Measurement accuracy |
|
|
R/S |
0°C ≦ TS ≦ 100°C |
±7.0°C |
||
|
100°C < TS ≦ 300°C |
±5.0°C |
|||
|
R:300°C < TS ≦ 1600°C |
±(0.05% of rdg +3.0°C) |
|||
|
S:300°C < TS ≦ 1760°C |
±(0.05% of rdg +3.0°C) |
|||
|
B |
400°C ≦ TS ≦ 600°C |
±5.5°C |
||
|
600°C < TS ≦ 1820°C |
±(0.05% of rdg +3.0°C) |
|||
|
K |
-200°C ≦ TS ≦ -100°C |
±(0.05% of rdg +3.0°C) |
||
|
-100°C < TS ≦ 1370°C |
±(0.05% of rdg +2.0°C) |
|||
|
E |
-200°C ≦ TS ≦ -100°C |
±(0.05% of rdg +3.0°C) |
||
|
-100°C < TS <span style="font-size:12.0pt;font-family:"Cambria Math","serif";mso-fareast-font-family: "Times New Roman";mso-bidi-font-family:"Cam |
||||