GESTER GT-RA08B-2 Protective Mask Leakage Tester

Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

Test aerosol: NaCl, Oil (Corn oil and paraffin oil)

Aerosol Concentration Detector:

Flame photometer (For NaCl aerosol test): 2sets

TSI Laser photometer (For oil aerosol test): 2sets

Oil aerosol generator:

Particle concentration: 20mg/m³~30mg/m³

Aerodynamics size distribution of particle: Approx 0.02~2um

Mass median diameter: Aapprox 0.3um

NaCl aerosol generator:

Particle concentration: 4mg/m³~12mg/m³

Aerodynamics size distribution of particle: Approx 0.02~2um

Mass median diameter: Aapprox 0.6um

Aerosol occurrence amount: ≥ 100L/min

Sampling pump adjust range: 0.50~4L/min

Particle sampling flow: 1~2L/min

Aerosol concentration detector dynamic range: 0.001~200mg/m³, accuracy 1%, response time ≤500ms

Power requirement: 1phase 220V 50Hz, 2KW

Air requirement: 0.5MPa, 120L/min

Dimension:

Outer size of testing room: 2000mm×1600mm×2600mm(L×W×H)

Control cabinet: 880mm×520mm×1400mm (L×W×H)

Standard accessories:

  • Flame photometer 2sets

  • TSI Laser photometer 2sets

  • NaCl aerosol generator 1set

  • Oil aerosol generator 1set

Datasheet