Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
CALIBRATION
TRACEABILITY 17025 Traceable to NIST
OPERATING AMBIENT CONDITIONS
Temperature : 18 °C to 28 °C
Humidity: Up to 80 % RH
STORAGE CONDITIONS
Temperature: -20 °C to 50 °C
Humidity: 0 % to 95 % RH (non-condensing)
Altitude: 2000 m
CONTROLLED RANGE
Temperature: 5°C to 55°C
Humidity @ 18°C : 5 % to 95 % RH
Humidity @ 35°C 5 % to 85 % RH
Humidity @ 23°C: 5 % to 95 % RH
Humidity @ 40°C 5 % to 75 % RH
Humidity @ 30°C: 5 % to 90 % RH
Humidity @ 50°C 5 % to 50 % RH
CALIBRATION SYSTEM ACCURACY
Temperature ± 0.12 °C Or Better Typically ± 0.10 °C
Humidity ± 1.00 % RH Or Better Typically ± 0.80 %
Controller type: PID Controller
UNIFORMITY
Temperature: 0.10 °C Or Better Typically ± 0.10 °C
Humidity: 0.30 % RH @ 18 °C to 28 °C Or Better Typically ± 0.25 %
STABILITY
Temperature: 0.05 °C
Humidity: 0.15 % RH @ 18 °C to 28 °C
RAMP / SOAK RATE OF CHANGE
Depends on ambient conditions
Temperature
High → Low 1.50 °C / minute (Typical)
Low → High 3.00 °C / minute (Typical)
Humidity
High → Low 5.00 % RH / minute (Typical)
Low → High 10.00 % RH / minute (Typical)
WARM-UP PERIOD
Ambient Condition 30 Minutes Maximum
Cold 30 Minutes
RESOLUTION
Display
Temperature: 0.01 °C
Humidity: 0.01 % RH
Dew Point: 0.01 °C (Calculated)
USB
Temperature: 0.01 °C
Humidity: 0.01 % RH