Range: Hot block: 50°C to 350°C (122°F to 662°F)
Cold block: –15°C to 110°C (5°F to 230°F), (–8°C [18°F] with the hot block at 350°C [662°F])
Accuracy: Hot block: ±0.6°C
Cold block: ±0.2°C
Stability: ±0.05°C
Well-to-well uniformity: ±0.1°C
Display resolution: 0.1°
Heating times: Hot block: 30 minutes from 25°C to 350°C
Cold block: 15 minutes from 25°C to 110°C
Cooling times: Hot block: 40 minutes from 350°C to 100°C
Cold block: 16 minutes from 25°C to –15°C
Stabilization times: 8 minutes
Well depth: 4 in (102 mm)
Computer interface: RS-232 included
Power: 115 V AC (±10 %), 3 A, or 230 V AC (±10 %), 2 A, specify, 50/60 Hz, 280 W
Size (H x W x D): 178 x 267 x 248 mm (7 x 10.5 x 9.75 in)
Weight: 4.5 kg (10 lb)