Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Liquid Lubricated Gas Operated
Pressure Range: 10 to 2,000 psi
Accuracy: ±0.015% of reading
Test port adaptors: 1/8, 1/4, 3/8 and ½ NPT or BSP
Instrument size : 17.5 W x 12 D x 8.5 H
Option:
- 5540 Metric and 5541 BSP Adapter Sets:
+ P5540 Metric: M14 x 1.5, M20 x 1.5, 1/8 and 1/4 BSP
+ P5541 BSP: 1/8, 1/4, 3/8 and 1/2 BSP
+ P5542 NPT: 1/8, 1/4, 3/8 and 1/2 NPT
- P5531 Dirt/Moisture Trap: Pressure range 0 to 500 psi (35 bar)
P5532: Dirt/Moisture Trap Pressure range 0 to 3,000 psi (210 bar)
P5543: Angle Adapter:
P5544: Two Gauge Stand
P5551 Pointer Remover/Punch
Accuracy |
|
| Accuracy* | ±0.015% of reading (±0.008% optional)
*Accuracy based on % of reading from 10% to 100% of the piston range when used in accordance with the corrections found on the calibration certificate. Below 10% ± (accuracy class) x 10% of the piston range. |
Mass and PCU Materials of Construction |
|
| Standard weight material | Series 3 non-magnetic austenitic stainless steel |
| Weight density | 7.8 g/cm³ |
| Optional fractional weights | Solution heat treated aluminum |
| Weight density | 2.7 g/cm³ |
| Models P3011, P3012, P3013, P3022, P3023, P3025 (V) | Piston Material: Ceramic Cylinder Material: Martensitic steel Coefficient of Expansion: 11 ppm/°C |
| Models P3014, P3015, P3025 (P) P3031, P3032 (P & V) | Piston Material: Tungsten carbide Cylinder Material: Martensitic steel Coefficient of Expansion: 16.5 ppm/°C |
| Model P3016 | Piston Material: Tungsten carbide Cylinder Material: Tungsten carbide Coefficient of Expansion: 11 ppm/°C |
General |
|
| Test port adaptors | 1/8, 1/4, 3/8 and ½ NPT or BSP |
| Instrument weight | 24 lbs |
| Instrument size | 17.5 W x 12 D x 8.5 H |
| Mass set weight (typical) | 65 lbs |
Minimum Standard Weight Increments |
|
| P3012, P3022 Pressure | 1 inH2O, (5 mbar) |
| P3013, P3023 Pressure | 2 inH2O, (10 mbar) |
| P3014, P3015, P3025 Pressure | 1 psi, (0.1bar) |
| P3016, P3031, P3032 | 1 psi, (0.1bar) |
| P3011, P3022, P3023, P3025 Vacuum | 0.2 inHg (10 mbar) |
Optional Fractional Weights |
|
| P3014, P3015, P3025 Pressure | 0.1 psi (0.01 bar) |
Built-in Hand Pumps |
|
| Pressure mode | Max pressure 300 psi (20 bar) |
| Vacuum mode | To 90% vacuum |
Operating Fluid for Liquid Lubricated Models |
|
| 55-655 | mineral oil (standard) |
| Krytox | for oxygen-safe applications (optional) |
Options |
|
| PressCal Software | Windows-based software program that allows users to easily apply all necessary corrections to enhance the deadweight tester performance. Calibration details are then stored and/or used to automatically create a calibration certificate. PressCal is provided as standard with all 0.008% instruments. |
| Krytox | As operating fluid for liquid lubricated gas operated models. |