Operating temperature: 10 °C to 40 °C (50 °F to 104 °F)
Storage temperature: -20 °C to 60 °C (-4 °F to 140 °F)
Humidity: 10 % to 90 % non-condensing
Display: LCD display
Communications: USB device port for computer control
Modes of operation: Manual and remote
Power: Internal rechargeable NiMH battery pack for nine hours (typical) operation after full charge or the battery charger can operate the analyzer and charge the battery simultaneously
Battery charger: 100 V to 240 V input, 15 V/1.5 A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle
Enclosure: ABS plastic housing
Dimensions (WxDxH): 32 cm x 24 cm x 13 cm (13 in x 9.5 in x 5 in)
Weight: 3.02 kg (6.6 lb, 0.1 oz)
Safety standards :
CE: IEC/EN61010-1 2nd Edition; Pollution degree 2
CAN/CSA-C22.2 No 61010-1; UL61010-1
C-Tick: Australian EMC
Electromagnetic compatibility standards (EMC): European EMC: EN61326-1
Load resistance:
Resistance: 50 Ω
Accuracy: ± 1 %, non-inductive (< 2 μH)
Pulse trigger level: 20 V
Pulse width
Range: 1 ms to 50 ms
Accuracy: ± 0.1 ms
Voltage
Range: 20 V to 5000 V
Accuracy: ± (1 % of reading + 2 V)
Current
Range: 0.4 A to 100 A
Accuracy: ± (1 % of reading + 0.1 A)
Tilt (biphasic and pulsed biphasic)
Range: 1 % to 99 %
Accuracy: ± 1 digit
Interphase delay (biphasic and pulsed biphasic)
Range: 0.1 ms to 9.9 ms
Accuracy ± 0.1 ms
Frequency (pulsed biphasic only)
Range: 2000 Hz to 8000 Hz
Accuracy: ± 1 % of reading
*For detailed specifications, please refer to the Datasheet.