115 VAC, 50/60 Hz
Temperature Control Method: Direct Heat and Air Jacket using Microcontroller PI
Ambient Temperature Range: 18 to 32˚C (64 to 89.6˚F)
Temperature Range, °C: Ambient +5 to 60
Temperature Uniformity, °C*: < ±0.5
Temperature Accuracy, °C*: < ±0.1
Temperature Recovery Time** (after 30 seconds door opening, 98% from initial value): ≤5 minutes
CO2 Control System: Microcontroller PI
CO2 Range, % CO2: 0-19.5
CO2 Accuracy, % CO2***: ±0.1
CO2 Sensor: Infrared (IR) Sensor
CO2 Recovery Time**** (after 30 seconds door opening, 98% from initial value): ≤5 minutes
Humidity
Humidification Method: Humidity pan
Humidity Range, %RH (at 37ºC)*****: 85-90
Physical Construction
Interior Volume: 170 L
External Dimensions (W x D x H): 660 x 660 x 900 mm
Internal Dimensions (W x D x H): 505 x 535 x 633 mm
Chamber Construction
Main Body: Electrogalvanized steel with Isocide™ antimicrobial coating
Interior Material: Stainless steel, type 304
Number of Shelves: 3
Maximum Number of Shelves: 7
Electrical Configuration: 110-130 VAC, 50/60 Hz
Full Load Amps: 9.2 A
Electrical Configuration: 220-240 VAC, 50/60 Hz
Full Load Amps: 4.2 A
Net Weight: 101 kg
All data recorded is specified for standard models with unloaded chambers and tested under optimum factory setting conditions of 23°C and 60% ambient humidity.
Contamination Control Methods
1) Main body is electro-galvanized steel with Isocide™ antimicrobial coating;
2) 90˚C moist heat OVERNIGHT decontamination cycle (HPA validated);
3) ULPA filter******
4) 0.2-micron inlet filter for gas inputs;
5) 1-micron air circulation filter
6) UV Lamp (for CLM-UV models only)
*Results are achieved when tested at 37°C as set point. Results may vary if set point changes and calibration is needed.
**For temperature not exceeding 37.2°C.
***Results are achieved when tested at 5% CO2 as set point. Results may vary if set point changes and calibration is needed.
****For CO2 level not exceeding 5.2%.
***** Esco does not guarantee condensation-free chamber at humidity level higher than 90%.
******Not available for 50 L models.