For full functionality of this site it is necessary to enable JavaScript.
EMIN.CO.TH
0
Product image

ESCO CCL-240B-8-HITEMP CelCulture® CO₂ Incubator (240 L, 230 VAC, 50/60 Hz)

ModelCCL-240B-8-HITEMP
P/N2170324
Contact
Secure Checkout
Quality Engagement
Easy change and return
Delivery Available

IR SENSOR MODEL WITH STAINLESS STEEL CHAMBER

CelCulture® Incubator 240L: IR Sensor, CO2 Control, ULPA, Moist Heat Decon, 230 VAC 50/60Hz


TEMPERATURE:

Temperature Control Method: Direct Heat and Air Jacket Using Microcontroller PI

Ambient Temperature Range*: 18 to 32°C (64.4 to 89.6°F)

Temperature Control Range, °C*: Ambient +5 to 60

Temperature Uniformity, °C*: <± 0.5

Temperature Fluctuation, °C*: <± 0.2

Temperature Recovery Time** (after 30 seconds door opening, 98% from initial value): ≤6 minutes


CO2:

CO2 Control System: Microcontroller PI

CO2 Control Range, % CO2: 0-19.5

CO2 Fluctuation, % CO2***: ± 0.2

CO2 Sensor: Infrared (IR) Sensor

CO2 Recovery Time**** (after 30 seconds door opening, 98% from initial value): Standard Unit: ≤5 minutes; Suppressed O2 model: ≤8 minutes


HUMIDITY:

Humidification Method: Humidity Pan

Humidity Range, %RH (at 37 °C)*******: 85-90


PHYSICAL CONSTRUCTION:

Interior Volume: 240 L

External Dimensions (W x D x H): 750 x 770 x 906 mm

Internal Dimensions (W x D x H): 595 x 640 x 633 mm

Chamber Construction: Main Body Electrogalvanized steel with Isocide™ antimicrobial coating

Interior Material: Stainless steel, type 304

Number of Shelves: 4

Maximum Number of Shelves: 7


Electrical Configuration 220-240 VAC, 50/60 Hz

Full Load Amps: 6.5 A


Net Weight: 121 kg


CONTAMINATION CONTROL

Contamination Control Methods

1) Main body is electro-galvanized steel with Isocide™ antimicrobial coating;

2) 90˚C moist heat OVERNIGHT decontamination cycle (HPA validated);

3) ULPA filter

4) 0.2-micron inlet filter for gas inputs;

5) 1-micron air circulation filter


All data recorded is specified for standard models with unloaded chambers and tested under optimum factory setting conditions of 23°C and 60% ambient humidity.

*Results are achieved when tested at 37°C as set point. Results may vary if set point changes and calibration is needed.

**For temperature not exceeding 37.2°C.

***Results are achieved when tested at 5% CO2 as set point. Results may vary if set point changes and calibration is needed.

****For CO2 level not exceeding 5.2%.

*****Results are achieved when tested at 5% O2 as set point. Results may vary if set point changes and calibration is needed.

****** For O2 level not lower than 4.8%.

******* Esco does not guarantee condensation-free chamber at humidity level higher than 90%.