Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

SUPPRESSED O2 MODEL WITH INTEGRATED COOLING SYSTEM
CelCulture® Incubator 170 L: IR Sensor, CO2 & O2 Control, Moist Heat Decon, Peltier System, 230 VAC, 50/60 Hz
O2 /T-SERIES (FOR SUPPRESSED O2 MODEL)
O2 Control System: Microcontroller PI
Range, % O2: 0-18
Accuracy*****, % O2: ±0.1
Sensor: Zirconia O2 sensor
TEMPERATURE:
Temperature Control Method: Direct heat & Air Jacket using Microcontroller PI
Ambient Temperature Range: 18 to 32°C (64.4 to 89.6°F)
Temperature Range, °C: 8°C below ambient to 60°C
Temperature Uniformity*, °C:
Heating mode: < ±0.7
Cooling mode: < ±2.0
Temperature Accuracy*, °C: < ±0.1
Temperature Recovery Time** (after 30 seconds door opening, 98% from initial value): ≤5 minutes
CO2:
CO2 Control System: Microcontroller PI
Range, % CO2: 0-19.5
Accuracy***, % CO2: ±0.1
Sensor: Infrared (IR) Sensor
Recovery Time**** (after 30 seconds door opening, 98% from initial value):
Standard Unit: ≤5 minutes
Suppressed O2 model: ≤6 minutes
HUMIDITY:
Humidity Method: Humidity Pan
Humidity Range, %RH (at 37°C)*******: 85-90
PHYSICAL CONSTRUCTION:
Interior Volume: 170 L
External Dimensions (W x D x H): 660 x 789 x 900 mm
Internal Dimensions (W x D x H): 505 x 535 x 633 mm
Chamber Construction: Main Body Electrogalvanized steel with Isocide™ antimicrobial coating
Interior Material: Stainless steel, type 304
Number of Shelves: 4
Maximum Number of Shelves: 7
ELECTRICAL CONFIGURATION:
110-130 VAC, 50/60 Hz: Full Load Amps: 9.2 A
220-240 VAC, 50/60 Hz: Full Load Amps: 4.2 A
Net Weight: 113 kg
CONTAMINATION CONTROL:
Contamination Control Methods:
1) Main body is electro-galvanized steel with Isocide™ antimicrobial coating;
2) 90˚C moist heat OVERNIGHT decontamination cycle (HPA validated);
3) ULPA filter********
4) 0.2-micron inlet filter for gas inputs;
5) 1-micron air circulation filter
All data recorded is specified for standard models with unloaded chambers and tested under optimum factory setting conditions of 23°C and 60% ambient humidity.
*Results are achieved when tested at 37°C as set point. Results may vary if set point changes and calibration is needed.
**For temperature not exceeding 37.2°C.
***Results are achieved when tested at 5% CO2 as set point. Results may vary if set point changes and calibration is needed.
****For CO2 level not exceeding 5.2%.
*****Results are achieved when tested at 5% O2 as set point. Results may vary if set point changes and calibration is needed.
****** For O2 level not lower than 4.8%.
******* Esco does not guarantee condensation-free chamber at humidity level higher than 90%.