Temperature Range
Ambient Temperature @ 23ºC: 33°C to 650 °C
Ambient Temperature @ 0ºC: 10°C to 650°C
Stability ( measured over 30 mins): ± 0.05 °C
Uniformity
Axial (lower 40 mm) : ± 0.35 °C
Radial: ± 0.15 °C
Accuracy
with Internal Reference Sensor: ± 0.30 °C
with External Reference Sensor SRS (*): ± 0.15°C ( TC version only)
Display Settings
Set Point Range: -0°C to 650 °C
Resolution: user-selectable 0.1 , 0.01 C
Units: user-selectable C, F,K
Calibration Volume
Immersion depth: 185 mm
Insert Diameter: 35 mm
Standard Insert Supplied: Ø3.5-5-6.5-8.5-12.5 mm pockets
Heating time: 33 to 650 ºC : 36 min
Cooling Time:
650 to 100 ºC : 39 min
100 to 50ºC : 19 min
50 to 33ºC: 21 min
Thermocouple Input (TC version)
Type: K,N,R,S,J,E
Range: 0 to 1700 °C
Accuracy: ± 1°C Include CJC error
Resolution: user-selectable 0.1 , 0.01 °C
RTD’S input ( TC version only)
Type: Pt 100 (3/4 wires) IEC
Range: -100 C to 700 °C
Accuracy: ± 0.3 °C
Resolution: user-selectable 0.1 , 0.01 °C
General
Computer interface: RS 232 as standard ( UBS adapter supplied as option) and EIUK software included
Switch Test: Yes With automatic log of switch open and switch closed
Ramp: Yes fully programmable
Weight: 10 Kg
Dimensions: 160x340x330 mm
Power Requirement: 115 or 230 V – 50/60 Hz
Power: 800 VA
Notes: * Require SRS sensor and special UKAS matching calibration. Long term drift not included
Scope of Supply: Each PTB600 is supplied with: Power Cable, RS233 cable, switch test lead, standard insert, insert removal tool, traceable calibration certificate and instruction manual