Dwyer IEFB-HB-PG-T10-LCD Thermal Energy Meter for 4-36" Pipes, LCD, Multi-Output, High Accuracy (± 1 % RDG, 250 °F, 1 in, BSPT Male, Aluminum, 10 in, 12 to 42 Vdc, 12 to 36 Vac, Cable Gland with Screw Terminal block)
Accuracy: ± 1 % RDG
Impedance: 4 to 20 mA: 536 Ω; 5V: 500 Ω; 10V: 1.27k Ω
Materials: Aluminum
Compliance: BTL
Conductivity: > 20 µS/cm
Display Note: 2 x 2" (50 x 50 mm) Graphic LCD with Backlight
Display Size: 71 mm (2.8 in)
Display Type: LCD
Accuracy Note: High accuracy 4 to 10˝ (100 to 250 mm) pipe; 1% of reading, High accuracy units: ± 0.5% of reading at calibrated velocity, ± 1% of reading from 2 to 20 ft/s (0.6 to 6 m/s) ± 0.02 ft/s (± 0.006 m/s) at < 2 ft/s (0.6 m/s); Standard accuracy units: ± 1% FS
Output Detail: (1) Analog: 4-20 mA, 0-5 V, 0-10 V or 2-10 V (display selectable); (1) Pulse/Frequency: 0-15 V peak pulse, 0 to 500 Hz or scalable pulse output (display selectable); (2) Alarm: (1) Empty pipe detection or minimum/maximum velocity, (display selectable); (1) Reverse flow output indication
Output Signal: Current, Voltage, DC Pulse
Control Output: Alarm
Data Interface: N/A
Pipe Size, Max: 10 in
Pipe Size, Min: 4 in
Supply Voltage: 12 to 42 Vdc, 12 to 36 Vac
Pressure Limits: Up to 400 psi
Resistance Note: Matched 4 Wire Platinum RTD’s
Enclosure Ratings: IP66, NEMA Type 4X
Materials, Wetted: Polymer, Silicone, Stainless Steel
Relative Humidity: 10 to 90% Non-Condensing
Supply Power Note: 12 to 42 Vdc, 0.25 A @ 24 Vdc; 12 to 36 Vac
Supply Power Type: AC/DC
Pressure Drop Note: < 0.1 psi at 12 ft/s in 4" (<0.01 bar at 3.7 m/s in 100 mm) and Larger Pipe
Pressure Drop, Max: 0.1 psi
Media Compatibility: Acid Solutions, Conductive Liquids, Saltwater
Electrical Connection: Cable Gland with Screw Terminal block
Linear Flow Rate, Max: 20 ft/s
Linear Flow Rate, Min: 0 ft/s
Process Connection Size: 1 in
Process Connection Type: BSPT Male
Temperature Probe Style: General Purpose Probe
Temperature Sensor Type: RTD
Process Temperature, Max: 250 °F
Process Temperature, Min: 15 °F
Storage Temperature, Max: 185 °F
Storage Temperature, Min: -40 °F
Temperature Compensation: 140 to 220°F < 2% error over ± 30°F Change , 40 to 70°F < 2% error over ± 10°F Change
Straight Run Requirements: >10 Diameter Upstream, >5 Diameter Downstream
External Probe Lead Length: 3 m (10 ft)
Operating Temperature, Max: 158 °F
Operating Temperature, Min: -4 °F