Measured value: Temperature, Relative humidity, Counted values (Dew point…)
Temperature sensor - external probe PT1000:
Measuring range: -200 to +600 °C
Accuracy: ±0.2 °C (-200 to +100 °C);
±0.2 % from reading from +100 to +600 °C
Resolution: 0.1 °C
Temperature + humidity sensor - external probe T+RH DIGI:
Measuring range: according to the connected probe Digi/M
Accuracy: according to the connected probe Digi/M
Resolution: 0.1 °C
Dew point:
Measuring range: according to the connected probe Digi/M
Accuracy: according to the connected probe Digi/M
Resolution: 0.1 °C
Temperature sensor - Thermocouple K, N:
Measuring range: -200 to +1300 °C
Accuracy: ± 0.3 % from the Measured value + 1.5 °C
Resolution: 0.1 °C
Temperature sensor - Thermocouple J:
Measuring range: -200 to +750 °C
Accuracy: ± 0.3 % from the Measured value + 1.5 °C
Resolution: 0.1 °C
Temperature sensor - Thermocouple S:
Measuring range: 0 to +1700 °C
Accuracy: ± 0.3 % from the Measured value + 1.5 °C
Resolution: 0.1 °C
Temperature sensor - Thermocouple B:
Measuring range: 100 to +1800 °C
Accuracy: ± 0,3 % from the Measured value + 1 °C in the range +300 to +1800 °C
Resolution: 0.1 °C
Temperature sensor - Thermocouple T:
Measuring range: -200 to +400 °C
Accuracy: ± 0.3 % from the Measured value + 1.5 °C
Resolution: 0.1 °C
DC voltage:
Measuring range: -60 to +140 mV; -18 to +18 mV
Accuracy: ±100 µV; ±20 µV
Resolution: 0.1 µV
Operating temperature: -10 to +60 °C
Channels: 4x external temperature probe
Counted values: Dew point, absolute humidity, specific humidity, mixing ratio, specific enthalpy
Memory: 1 000 000 values (noncyclic record), 600 000 values (cyclic record)
Recording interval: 1s to 24 hours
Display and alarm refresh: each 10 s
Recording mode: noncyclic - data logging stops after filling the memory
cyclic - after filling memory oldest data is overwritten by new
Available temperature units: degrees Celsius, degrees Fahrenheit
Communication interface: Ethernet, USB, RS232
Ethernet communication protocols: WWW, SNMPv1, XML, DATALINK
Alarm protocols E-Mail
Power: AC adapter 230Vac/5Vdc (neccessary for Ethernet interface)
Battery life: 3 - 4 weeks
Protection class: IP20
Dimensions (without connectors): 178 x 95 x 37mm, without attached cables
Weight (including batteries): approx. 380 g
Included in delivery:
M1220
Instruction Manual
Traceable Calibration Certificate
3x Rechargeable Battery AA
Power Supply
USB Cable
Wall Holder
FREE FULL VERSION Analytical Software COMET Vision