Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Detection limit
Benzene: ~10 ppt with pre-concentration trap
Toluene, Ethylbenzene, m+p-Xylenes, & O-xylene: <50 ppt
Configuration & cycle time: BTEX: 15 min
Detection range
0-15 ppb
0-100 ppb with other PID lamp
Measurement
Detector: PID
Temporal resolution: 0.1 seconds
Response time: One measurement every 10 minutes (default settings)
Analysis sample condition: Gas T°: 5 - 40°C; Gas RH: 20 - 90%; Atmospheric pressure
Calibration: Gaseous BTEX mixture or gaseous Benzene
Sampling
200μL loop or carbon trap
Gas flow rate: between 10 to 100 mL min-1
Carrier gas: Nitrogen 4 bar inlet pressure and 2.5 mL min-1
Supply inlet connection: 1/8’’
Instrument supply
Power supply: Input 100 - 240V ±10%; 1.5 A max; 47 - 63 Hz - Output 15V; 6.67A 100W
Autonomy on battery: Up to 4h
Power consumption: max 75 W
General
Dimensions (analyzer): 32×28×15 cm; 6.0 kg
Dimensions (suitcase): 56 x 45 x 25 cm; 18 kg
Operational conditions: 0 - 40°C / 20 - 80% RH
Storage conditions: -20°C - +40°C / 0 - 85% RH
Display: 7” TFT display; resolution 800 x 480; integrated touchscreen