airmoVOC C6-C12 XPERT
C6 to C12 Analysis
Range: C6 (Hexane, Dimethylbutane) to C10 (Diethylbenzene, Naphthalene) and Dodecane
Detection Limit:
Tri-methylbenzene: 10 ppt or 0.05 µg/m³
Benzene: 1 ppt with airmoVOC Expert
Detection Range: 0.05 to 45 µg/m³ for Tri-methylbenzene and Benzene in
Note: Be careful with total mass and mass by compounds.
Amplification 1: On demand to decrease the sampling volume.
Relative Standard Deviation (RSD):
< 0.3 % over 48 h (Retention Time – RT)
< 3 % over 48 h (Concentration – Conc.)
Results:
Data storage
MODBUS / JBUS communication protocol (option)
Cycle Time: 15 min, 30 min, 40 min, 60 min
Gas Supply:
H₂ (FID and carrier gas): 30 mL/min (inlet 2 bars; 1/16’’ Swagelok)
Air (FID): 180 mL/min (inlet 3 bars; 1/8’’ Swagelok)
Sample inlet (vacuum pump): 1/4’’ Swagelok
Pneumatic valve: 90 mL/commutation
Sample Volume: 30 to 700 mL (programmable) (Sampling time up to 40 minutes)
Power Supply:
Main: 230 V / 115 V or 50 Hz / 60 Hz
Battery: 24 V (option)
Electrical Consumption: Mean: 150 VA; Peak: 360 VA
Dimensions and Weight:
Rack: 19’’ (5U)
Height: 222 mm
Width: 482 mm
Depth: 600 mm
Net Weight: 22 kg