Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

MEASUREMENT PARAMETERS
> Conductivity
Measurement range: 0.050 µS/cm to 500.0 mS/cm
Resolution (R): 0.001 to 0.1 (depending on range)
Intrinsic uncertainty of the instrument (without the sensor): ± 0.5% reading ± R
> TDS
Measurement range: 0.001 mg/l to 499.9 g/l
Resolution (R): 0.001 to 0.1 (depending on range)
Intrinsic uncertainty of the instrument (without the sensor): ± 0.5% reading ± R
> Resistivity
Measurement range: 2.000 Ω.cm to 19.99 MΩ.cm
Resolution (R): 0.001 to 0.01 (depending on range)
Intrinsic uncertainty of the instrument (without the sensor): ± 0.5% reading ± R
> Salinity
Measurement range: 2.0 to 42.0 psu
Resolution (R): 0.1
Intrinsic uncertainty of the instrument (without the sensor): ± 0.5% reading ± R
> Temperature
Measurement range (instrument only): - 10.0 to + 120.0 °C (14.0 to 248.0 °F)
Resolution (R): 0.1 °C (0.1 °F)
Intrinsic uncertainty of the instrument (without the sensor): < 0.4 °C (<0.7°F)
Reference temperature available: 20/25 °C (68/77 °F)
Calibration: 1 point
Temperature compensation: Automatic (ATC) or manual (MTC) – Linear, non-linear
Conductivity cell: XCP4ST1 (supplied), 4-pole conductivity sensor with built-in Pt1000 temperature sensor (0...100 °C), 8-pin DIN connector, 1 m cable
> Data storage
Date and time: Yes
Memory: > 100,000 measurements
> Connectors
Sensor input: 8-pin DIN (adapters for BNC, S7 and Jack available as options)
Communication interface: Type B micro-USB (USB device) 12 Mbit/s
Batteries / Battery life: 4 x 1.5 V AA or LR6 alkaline batteries, approx. 300 hours of continuous operation – Automatic shutdown after 3, 10 or 15 min without activity (adjustable)
IP rating: IP67
> Environmental conditions
Storage range (without battery): - 20 to + 70 °C
Operating range: - 10 to + 55 °C
Dimensions (with sheath): 211 x 127 x 54 mm
Weight (without electrode): 600 g
CA 10141 conductivity meter delivered in a site-proof case with 1 x XCP4ST1 conductivity sensor with built-in temperature sensor, 4 AA or LR6 alkaline batteries, 1 protective sheath mounted on the instrument, 1 ready-to-use standard reference solution (NIST-compliant) with conductivity of 1408 μS/cm, 1 plastic beaker, 1 USB - micro USB cable, 1 wrist strap, quick start-up guides (one per language).