Measuring Range
Rangeability: 10:1 (most sizes)
Metering Tube
Standard: 316/316L (dual certified stainless steel)
Flanges and End Fittings
Standard: 316/316L (dual certified stainless steel)
Accuracy: 5%, VDI/VDE class 6
Repeatability: 0.25% Full Scale
Scale type/material: Dark increments with white background/Aluminum
Installation orientation and location: Vertical (within 5% of true vertical), bottom inlet, top outlet. Do not locate in proximity of other magnetic interfering components
Connections
Flanged: Weldneck flanges
Equivalent to ANSI B16.5: ANSI 1/2” to 2” 150# RF to 300# RF
DIN 2527/ EN 1092-1: DIN PN 40
Flange finish: 3.2 - 6.3 Ra
Threaded female: 1/2” to 2” NPT Female
O-ring material
Threaded female std: Viton or Teflon
Floats
Standard: 316L stainless steel
Protection Category
Indicator only: IP66/67 NEMA 4X
Transmitter ALU: IP66/67 NEMA 4X**
Transmitter SS: IP66/67 NEMA 4X
Indicator Housing & Cover material
Indicator only ALU: Die cast Aluminum (Alloy 380), epoxy paint, glass window
Transm/Alarm/HiTemp ALU: Die cast Aluminum (Alloy 380), epoxy paint, glass window
Indicator only SS: Cast 316L stainless steel, glass window
Transm/Alarm/HiTemp SS: Cast 316L stainless steel, 316 stainless steel hardware, glass window
Pressure/Temperature
Maximum Fluid Temperature: 300°C/570°F
Meter Dimensions
Needle Control Valves & Flow Controllers: Valves Sizes 7 - 12 / FCA Sizes 7.8