General:

Input voltage: 100 – 260 V, 50/60 Hz

Power consumption: max. 300 VA

Reverse voltage protection: up to 13 kV

Degree of protection: IP54 (in closed state)

Data interface: USB 2.0

Dimensions (W x H x D): 438 x 456 x 220 mm

Weight (incl. HV connection cable): Approx. 22 kg

Ambient temperature (operational): -10°C to +50°C

Storage temperature: -20°C to +60°C

Safety and EMC: CE-compliant in accordance with Low Voltage Directive (2014/35/EU), EMC Directive (2014/30/EU), EN 60068-2-ff Environmental testing

User interface available in 13 languages: Czech, Chinese (CN), Chinese (TW), Dutch, English, French, German, Italian, Korean, Polish, Portuguese, Russian, Spanish

Output voltage:

Frequency range: 0.01 – 0.1 Hz

VLF truesinus®: 1 – 26 kVrms (36 kVpeak)

VLF square wave voltage: 1 – 34 kV

DC voltage: ±1 – 34 kV

Resolution: 0.1 kV

Accuracy: 1%

Load range (VLF testing): 1 nF – 8 µF

Output current:

Measurement range: 0 – 14 mA

Resolution: 1 μA

Accuracy: 1%

Max. capacitive load:

0.5 µF at 0.1 Hz, 24 kVrms / 34 kVpeak

1 µF at 0.05 Hz, 24 kVrms / 34 kVpeak

8 µF at 0.01 Hz, 18 kVrms / 25 kVpeak

Dissipation factor measurement (frida TD):

VLF truesinus®: 1 – 26 kVrms

Load range: 10 nF - 8 µF

Resolution: 1 x 10^-6

Accuracy: 1 x 10^-4

Measurement range: 1 x 10^-4 – 21,000 x 10^-3

tan δ measuring frequency: 0.1 Hz

Automatic detection and compensation of leakage currents

With VSE box (optional)

BAUR Software 4 for office PC (office installation)

Used to evaluate test and measurement logs

frida TD standard delivery

▪ frida TD VLF tester and diagnostics device, incl.

– HV connection cable, 5 m (fix mounted)

– GDR 40-136 discharge and earth rod

– Earth cable, 5 m, with earth terminal

– Jumper plug for external emergency off unit

– BAUR Software 4 for office PC (office installation)

– HV connection set

– tan delta kit

– Mains supply cord, 2.5 m

– User manual

– Additional manual for dissipation factor measurement

– Pocket guide