Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts
Get exclusive volume discounts, bulk pricing updates, and new product alerts delivered directly to your inbox.
By subscribing, you agree to our Terms of Service and Privacy Policy.
Direct access to our certified experts

Measurement principle:
Vacuum Volumetric Physisorption
N2, Ar, Kr, CO2, H2, and other non-corrosive gases: Included
CH4, C2H6, C3H8, other alkanes, C6H6, C8H10: Included:
NH3, C5H5N, C3H6O, and other reactive gases: N/A
Pressure:
Range: 2x10-5 to 1100 Torr (2.6x10-8 to 0.997 p/p0 for N2 77K)
Resolution (MP): 2x10-5 Torr (2.6x10-8 p/p0 for N2 77K)
Resolution (XR): 1x10-6 Torr (1.3x10-9 p/p0 for N2 77K)
BET surface area:
Absolute detection limit: 0.1 m2 (N2 77K)
Specific detection limit: 0.01 m2/g (N2 77K)
Typical reproducibility: 1 % (measured on BAM P115)
Reproducibility limit: 2 % with 2 m2 in the cell
Pore size:
Range: 0.35 nm to 500 nm (diameter)
Typical reproducibility: 0.5 % (measured on BAM P115)
Dimensions (W x D x H): 79.5 cm x 70.1 cm x 107.9 cm (31.3 in x 27.6 in x 42.5 in) dependent on instrument configuration
Weight: 136.4 kg (300 lbs) dependent on instrument configuration
Operating environment:
Temperature: 15 °C to 35 °C (59 °F to 95 °F)
Humidity: 20% RH to 80% RH, non-condensing
Gas distribution manifold elastomers: FKM
Thermal conductivity detector (TCD): N/A
Gas ports:
7 ports standard (5 analysis, 1 helium, 1 degas/backfill)
7 additional analysis ports optional
Gas supply:
Purity: 99.999 %
Input pressure: 8 psig to 10 psig (0.55 bar to 0.69 bar)
Air: Input pressure: 50 psig to 100 psig (3.45 bar to 6.9 bar)
Electrical:
Supply: 100 to 240 VAC ±10%, 50 or 60Hz
Maximum consumption: 1600W (dependent on instrument configuration)