- Temperature Range:
650 ~ 1200°C (1202 ~ 2192°F) (MB 18)
800 ~ 2500°C (1472 ~ 4532°F) (MB 25)
- 3 Optics:
a = 600mm, 1000mm, 1500mm
- Spectral Ranges: 0.8 ~ 1.1 μm
- Resolution: 0.1°C
- Measurement Uncertainty (e = 1, t90 = 1 s, Tamb. = 25°C): 0.5% of reading in °C + 1°C
- Repeatability (e = 1, t90 = 1 s, Tamb. = 25°C): 0.1% of reading in °C + 1°C
- Emissivity ε: 0.05 ~ 1.0
- Sighting: LED targeting light
Electrical:
- Power Supply: 24 VDC ± 25%, ripple 500 mV; LED targeting light: 5 ~ 30 VDC, 35 mA
- Power Consumption: Max 0.6 W (without LED targeting light)
- Load: Max 700 Ω @ 24 V (max 100 Ω @ 12 V)
Environmental
- Protection system: IP65 (according to DIN 40 050)
- Ambient Temperature: 0 ~ 70°C (32 ~ 158°F)
- Storage Temperature: -20 ~ 70°C (-4 ~ 158°F)
- Weight: ≈450 g (≈0.992 lb)
- CE Label: According to EU directives about electromagnetical immunity
Communication and interface:
- Analog Output: 4 ~ 20 mA, linear
- Response Time t90: 20 ms ~ 10 s
- Scope of Delivery: Instrument with selected optic, works certificate, operation manual.